FDA Medical Device Recalls (openFDA)
Abbott Laboratories Diagnostic Div: Clinical Chemistry Multiconstituent Calibrator--- List number: lE65-02
- Recall number
- Z-0516-06
- Recalling firm
- Abbott Laboratories Diagnostic Div
- Product
- Clinical Chemistry Multiconstituent Calibrator--- List number: lE65-02
- Status
- Terminated
- Initiated
- 2005-12-23
- Posted
- 2006-02-14
- Terminated
- 2012-04-24
- Reason
- The two Multiconstituent Calibrator (MCC) lot numbers 19906M200 and 20879M100 failed to meet product performance expectations throughout the labeled expiration period, as a result of the shortened expiration dating, two lots are now expired. The expiration date for Lot number 26674M100 will be revised from 2006-07-31 to 2006-01-31.
- Root cause
- Other
- Action
- A Device Correction letter identifying the changes being made to the expiration dating for lot numbers 19906M200, 20879MI00 and 26674M1O0 along with a Customer Reply form will be distributed via FedEx to all AEROSET and ARCHITECT c8000 customers which utilize the Clincial Chemistry Multiconstituent Calibrator on December 27,2005. Outside the U.S. each country organization will need to determine the appropriate method for the effectiveness check.
- Quantity
- 13961
- Distribution
- Nationwide plus worldwide in Mexico,Venezuela, Chile, Brazil, San Salvador, Singapore, Korea, Trinidad, Grand Cayman, Guatemala, Columbia, Uruguay, Puerto Rico, Canada, Hong Kong, Australia, Bermuda, Jamaica, Costa Rica, Ecuador, Argentina, Dominican ReD., Germanv, Thailand, New Zealand, Bahamas, Panama, & Paraguay.
- Lot, serial or model codes
- Lot number(s): 19906M200 (expiration date 02/28/2006), 20879MIOO (expiration date 02/28/2006), & 26674MIOO (expiration date 07/31/2006)
- 510(k) numbers
- K981706
- PMA numbers
- not on file
- Product code
- JIX
- Firm FEI
- 2018433
- Firm city
- South Pasadena
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28