Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc: ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756.

Recall number
Z-0515-2018
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756.
Status
Terminated
Initiated
2017-06-07
Posted
not on file
Terminated
2019-05-08
Reason
Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots ending in 21 (C3021) as compared to Calibrator 30 kit lots ending in 20 (C3020).
Root cause
Under Investigation by firm
Action
An Urgent Medical Device Recall letter dated June 7, 2017 was distributed to all customers in the United States and an Urgent Field Safety Notice was distributed to all customers outside the United States who have received ADVIA Centaur Systems Calibrator 30 kit lot ending in 21(C3021). These letters instruct customers who operate the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems to discontinue use of C3021 . The letters inform customers that Calibrator 30 kit lots ending in 20 (C3020) and 23 (C3023) are not impacted and can be used. The letters inform customers who operate ADVIA Centaur CP systems to continue using their inventory of C3021, as they are not impacted by this issue. All affected domestic customers were sent a hard copy of the Urgent Medical Device Recall via Federal Express. Fax-back forms will be used to confirm that the customers received the Urgent Medical Device Recall letter.
Quantity
5992 units (US and OUS)
Distribution
US and OUS.
Lot, serial or model codes
Lot Numbers: 00387A21, 16673A21, 36243A21;  00388A21, 15448A21, 51211A21;  20640A21; 51024A21.  Expiration Date: 2018-03-08
510(k) numbers
K100293
PMA numbers
not on file
Product code
CHP
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register