FDA Medical Device Recalls (openFDA)
GE Healthcare: The LightSpeed RT16 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications. The system is capable of generating images for the guidance of minimally invasive procedures such as biopsy and ablation of tumors and pathology. The system allows imaging of obese patients, up to and including the obese population (BMI > 40). When used in the LightSpeed RT 16 configuration, the system acquires CT anatomical images that are clinically useful in the simulation and planning of radiation therapy for the treatment of cancer.
- Recall number
- Z-0506-2007
- Recalling firm
- GE Healthcare
- Product
- The LightSpeed RT16 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications. The system is capable of generating images for the guidance of minimally invasive procedures such as biopsy and ablation of tumors and pathology. The system allows imaging of obese patients, up to and including the obese population (BMI > 40). When used in the LightSpeed RT 16 configuration, the system acquires CT anatomical images that are clinically useful in the simulation and planning of radiation therapy for the treatment of cancer.
- Status
- Terminated
- Initiated
- 2006-12-08
- Posted
- 2007-02-15
- Terminated
- 2007-11-30
- Reason
- When the manual cradle release feature on the table is used to move a patient out of the gantry, a discrepancy of up to 6mm between the numerical display on the gantry and the actual position of the table could be introduced.
- Root cause
- Other
- Action
- A Product Safety Advisory Notice, dated December 8, 2006, was issued to all customers informing them of the safety issue and production correction. Customers are advised to operate the CT cradle as intended with the system''s electronic controls. If a critical need arises to release the cradle during a radiation therapy simulation scan, they are advised to independently verify patient position prior to marking the patient for therapy to avoid the potential inaccuracy. A software update will be installed to correct this problem by a GE representative.
- Quantity
- 22 units
- Distribution
- Worldwide, including USA, Puerto Rico, Albania, Argentina, Austria, Australia, Belarus, Belgium, Bulgaria, Brazil, Canada, Chile, China, Columbia, Denmark, Finland, France, Germany, Great Britain, Greece, Guam, Hong Kong, Honduras, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Malaysia , Martinique, Mexico, Morocco, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Reunion Island, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain , Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Venezuela, Vietnam, and Yugoslavia.
- Lot, serial or model codes
- All serial numbers
- 510(k) numbers
- ["K060052"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28