FDA Medical Device Recalls (openFDA)
ELEKTA SOLUTIONS AB: MOSAIQ Oncology Information System Software Builds 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 Software Version and UDI Codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201500026(10)3.1.3.1 3.1.3.2 (01)07340201500026(10)3.1.3.2 3.1.3.3 (01)07340201500026(10)3.1.3.3 3.1.3.4 (01)07340201500026(10)3.1.3.4 3.2.1.0 (01)07340201500071(10)3.2.1.0 3.2.1.1 (01)07340201500071(10)3.2.1.1 3.2.1.2 (01)07340201500071(10)3.2.1.2 3.2.1.3 (01)07340201500071(10)3.2.1.3 3.2.1.4 (01)07340201500071(10)3.2.1.4 3.2.2.0 (01)07340201500071(10)3.2.2.0 3.2.2.1 (01)07340201500071(10)3.2.2.1 3.2.3.0 (01)07340201500071(10)3.2.3.0 MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.
- Recall number
- Z-0503-2026
- Recalling firm
- ELEKTA SOLUTIONS AB
- Product
- MOSAIQ Oncology Information System Software Builds 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 Software Version and UDI Codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201500026(10)3.1.3.1 3.1.3.2 (01)07340201500026(10)3.1.3.2 3.1.3.3 (01)07340201500026(10)3.1.3.3 3.1.3.4 (01)07340201500026(10)3.1.3.4 3.2.1.0 (01)07340201500071(10)3.2.1.0 3.2.1.1 (01)07340201500071(10)3.2.1.1 3.2.1.2 (01)07340201500071(10)3.2.1.2 3.2.1.3 (01)07340201500071(10)3.2.1.3 3.2.1.4 (01)07340201500071(10)3.2.1.4 3.2.2.0 (01)07340201500071(10)3.2.2.0 3.2.2.1 (01)07340201500071(10)3.2.2.1 3.2.3.0 (01)07340201500071(10)3.2.3.0 MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.
- Status
- Open, Classified
- Initiated
- 2025-09-17
- Posted
- 2025-11-14
- Terminated
- not on file
- Reason
- When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule.
- Root cause
- Software design
- Action
- On 09/17/2025, the firm emailed an "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" to customers informing them when appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule. Customers are instructed to: When appending cycles for care plans that contain additional medication orders with a wave schedule, review and adjust the appended care plan cycles to reflect the clinical intent. Alternatively, wave orders can be removed from the care plan folder. Additional information can be found in: Customers with v3.1.3 - section 3.4.1 - page 35-36 in doc E062133 / 01 MOSAIQ 3.1.3.0 Instructions for Use Customers with v3.2.1 - section 3.4.1 - page 35-36 in doc E065117 / 03 MOSAIQ 3.2.1.0 Instructions for Use Customers with v3.2.2 - section 3.4.1 - page 35-36 in doc E067652 / 04 MOSAIQ 3.2.2.0 Instructions for Use Customers with v3.2.3.0 - section 3.4.1 - page 35-36 in doc E071967 / 01 MOSAIQ 3.2.3.0 Instructions for Use For questions - contact Quality Assurance at +46 70-962-0862 or email mats.premfors@elekta.com
- Quantity
- 4 systems that contain the affected software.
- Distribution
- Worldwide - U.S. Nationwide distribution in the state of NM and the countries of Canada, Malta, and South Africa.
- Lot, serial or model codes
- Software Versions: 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 UDI codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201500026(10)3.1.3.1 3.1.3.2 (01)07340201500026(10)3.1.3.2 3.1.3.3 (01)07340201500026(10)3.1.3.3 3.1.3.4 (01)07340201500026(10)3.1.3.4 3.2.1.0 (01)07340201500071(10)3.2.1.0 3.2.1.1 (01)07340201500071(10)3.2.1.1 3.2.1.2 (01)07340201500071(10)3.2.1.2 3.2.1.3 (01)07340201500071(10)3.2.1.3 3.2.1.4 (01)07340201500071(10)3.2.1.4 3.2.2.0 (01)07340201500071(10)3.2.2.0 3.2.2.1 (01)07340201500071(10)3.2.2.1 3.2.3.0 (01)07340201500071(10)3.2.3.0
- 510(k) numbers
- ["K223229"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 3015232217
- Firm city
- Stockholm
- Firm state
- not on file
- Firm country
- Sweden
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28