FDA Medical Device Recalls (openFDA)
Snug Seat Inc: Crocodile (size 3) are two-armed, wheeled walking aids, used by children with disabilities. Product Usage: The Crocodile is a walking aid for children with disabilities, who need supplementary assistance with their daily walking. The Crocodile is available in three sizes and can be used for children between the ages of 2 to 18 years.
- Recall number
- Z-0503-2015
- Recalling firm
- Snug Seat Inc
- Product
- Crocodile (size 3) are two-armed, wheeled walking aids, used by children with disabilities. Product Usage: The Crocodile is a walking aid for children with disabilities, who need supplementary assistance with their daily walking. The Crocodile is available in three sizes and can be used for children between the ages of 2 to 18 years.
- Status
- Terminated
- Initiated
- 2013-08-26
- Posted
- 2014-12-16
- Terminated
- 2017-12-04
- Reason
- The bolts, which secure the handles, on the Crocodile (size 3) could break, causing the handle to dislodge.
- Root cause
- Nonconforming Material/Component
- Action
- An Urgent Field Safety Notice letter dated August 26, 2013 and September 17, 2013 was sent to customers. The letter identified the affected product, problem and actions to be taken. New replacement bolts were provided to customers.
- Quantity
- 726 units
- Distribution
- Worldwide Distribution - U.S. Nationwide in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, LA, MA, MD, ME, MI, MN, MS, MT, NC, ND, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA & WV; International: Australia, Canada, China, Denmark, Estonia, Finland, France, Germany, Greece, Italy, Japan, New Zealand, Poland, Russia, Spain, The Netherlands, and The UK
- Lot, serial or model codes
- GMDN code; 31119
- 510(k) numbers
- ["K934560"]
- PMA numbers
- not on file
- Product code
- IOR
- Firm FEI
- 1053060
- Firm city
- Matthews
- Firm state
- NC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28