FDA Medical Device Recalls (openFDA)

COVIDIEN LLC: Puritan Bennett 980 (PB980) ventilator series: Puritan Bennett" 980 Pediatric Adult Ventilator Puritan Bennett" 980 Neonatal Ventilator Puritan Bennett" 980 Universal Ventilator Product Usage: The Puritan Bennett 980 Series Ventilator is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb). It is suitable for service in hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilatory support using medical oxygen and compressed medical air from either an internal air compressor or external air sources to deliver oxygen concentrations of 21% to 100%. Ventilatory support can be delivered invasively or non-invasively, to patients who require the followingtypes of ventilator support: Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or non-invasively (via mask or nasal prongs). Assist/ Control, SIMV or Spontaneous modes of ventilation

Recall number
Z-0497-2019
Recalling firm
COVIDIEN LLC
Product
Puritan Bennett 980 (PB980) ventilator series: Puritan Bennett" 980 Pediatric Adult Ventilator Puritan Bennett" 980 Neonatal Ventilator Puritan Bennett" 980 Universal Ventilator Product Usage: The Puritan Bennett 980 Series Ventilator is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb). It is suitable for service in hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilatory support using medical oxygen and compressed medical air from either an internal air compressor or external air sources to deliver oxygen concentrations of 21% to 100%. Ventilatory support can be delivered invasively or non-invasively, to patients who require the followingtypes of ventilator support: Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or non-invasively (via mask or nasal prongs). Assist/ Control, SIMV or Spontaneous modes of ventilation
Status
Terminated
Initiated
2018-09-18
Posted
2018-12-18
Terminated
2021-02-01
Reason
Software Update: External USB Drive performance and its impact on Graphic User Interface (GUI) functionality and labeling of the scalar waveform displayed on GUI during ventilation
Root cause
Device Design
Action
Medtronic notified U.S. consignees notified via Federal Express or certified mail on September 19, 2018, and the letter informs customers that Medtronic has developed a software update, Maintenance Release (MR) 5.4. Medtronic Service Engineers will load the MR5.4 software during the next scheduled or planned maintenance visit. Immediately notify all care environments in which the PB980 ventilators are used about this notification. " If your facility has distributed PB980 ventilators to other persons or facilities, please promptly forward a copy of this letter to those recipients. " Complete the attached form and return it as directed to confirm your receipt and understanding of this information. Any incidents related to this issue or if you have any questions, please contact our Technical Support Department immediately at 1-800-255-6774. Out--Side US accounts notified by Medtronic Regulatory Partners worldwide on 9/17/18 for translation in their respective country.
Quantity
9,593
Distribution
Worldwide Distribution - US Nationwide Foreign: Canada Argentina Chile Japan Hong Kong Taiwan Uruguay Austria Austria Azerbaijan Belgium Bosnia And Herzegovina Bulgaria Croatia Czech Republic France Germany Greece Hungary Ireland Israel Italy Kazakhstan Kenya Kuwait Latvia Lebanon Lithuania Netherlands Oman Pakistan Poland Portugal Romania Russian Federation Saudi Arabia Slovakia Slovenia South Africa Spain Sweden Switzerland Tunisia Turkey United Arab Emirates Iran United Kingdom
Lot, serial or model codes
Serial Numbers: 35B1300001 through 35B1300051  35B1400003 through 35B1402216   35B1500001 through 35B1502227  35B1600001 through 35B1601408  35B1700001 through 35B1702519  35B1800001 through 35B1801541  35B1400001  35B1601182, 35B1601183, 35B1601193, 35B1601197, 35B1601242 35B1601243, 35B1601245, 35B1702479, 35B1601001, 35B1601089
510(k) numbers
["K162738"]
PMA numbers
not on file
Product code
CBK
Firm FEI
1282497
Firm city
Mansfield
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register