FDA Medical Device Recalls (openFDA)
Invivo Corporation: Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388
- Recall number
- Z-0462-2019
- Recalling firm
- Invivo Corporation
- Product
- Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388
- Status
- Completed
- Initiated
- 2018-09-24
- Posted
- not on file
- Terminated
- not on file
- Reason
- The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard
- Root cause
- Software Design Change
- Action
- Customers were notified on about 09/20/2018 via an "URGENT - Medical Device Correction" letter. An initial letter was sent to distributors with instructions to provide the Field Safety Notice to all customers the devices were further distributed to. The "URGENT - Medical Device Correction" letter provided guidance on how to identify the Service Number and serial number on the Flex Cardio device, which is on the bottom right corner of the back of the device. Actions to be taken by the customer/user until Philips contacts the customer to arrange for a software update to address this issue, is if the IBP numeric values become frozen and do not update in real-time, opening the Waveform Setup Dialog and then immediately pressing the OK button will restore the proper display of the IBP numeric values. Alternative methods of performing the functions listed in the Problem Description section that do not cause the problem to occur include the following: 1. Change an IBP Site Label using the Pressure Setup Dialog or a macro; 2. Activate or deactivate the mean pressure waveform display option (On/Off) by using the Waveform Setup Dialog, the Pressure Setup Dialog or a macro; 3. Activate or deactivate the waveform hide option (On/Off) using the Waveform Setup Dialog, the Pressure Setup Dialog, the keyboard or a macro; 4. Activate or deactivate the pressure waveform filter option (On/Off) using a macro. A Philips representative will contact each customer regarding their affected Flex Cardio devices. A software update will be required on both the affected Flex Cardio device and on the host workstations driver. These actions will be implemented free of charge by Philips.
- Quantity
- 76 units
- Distribution
- Distributed to the Netherlands
- Lot, serial or model codes
- Model Number 453564669081; System Serial Number (UDI): US1814000001 ((01)00884838091474(21)US1814000001), US1814000002 ((01)00884838091474(21)US1814000002), US1814000003 ((01)00884838091474(21)US1814000003), US1814000004 ((01)00884838091474(21)US1814000004), US1814000006 ((01)00884838091474(21)US1814000006), US1814000007 ((01)00884838091474(21)US1814000007), US1814000008 ((01)00884838091474(21)US1814000008), US1814000011 ((01)00884838091474(21)US1814000011), US1814000012 ((01)00884838091474(21)US1814000012), US1814000013 ((01)00884838091474(21)US1814000013), US1814000014 ((01)00884838091474(21)US1814000014), US1814000016 ((01)00884838091474(21)US1814000016), US1814000017 ((01)00884838091474(21)US1814000017), US1814000018 ((01)00884838091474(21)US1814000018), US1814000019 ((01)00884838091474(21)US1814000019), US1814000020 ((01)00884838091474(21)US1814000020), US1814000022 ((01)00884838091474(21)US1814000022), US1814000023 ((01)00884838091474(21)US1814000023), US1814000024 ((01)00884838091474(21)US1814000024), US1814000027 ((01)00884838091474(21)US1814000027), US1814000028 ((01)00884838091474(21)US1814000028), US1814000029 ((01)00884838091474(21)US1814000029), US1814000030 ((01)00884838091474(21)US1814000030), US1814000031 ((01)00884838091474(21)US1814000031), US1814000032 ((01)00884838091474(21)US1814000032), US1814000033 ((01)00884838091474(21)US1814000033), US1814000034 ((01)00884838091474(21)US1814000034), US1814000035 ((01)00884838091474(21)US1814000035), US1814000036 ((01)00884838091474(21)US1814000036), US1814000037 ((01)00884838091474(21)US1814000037), US1814000038 ((01)00884838091474(21)US1814000038), US1814000039 ((01)00884838091474(21)US1814000039), US1814000040 ((01)00884838091474(21)US1814000040), US1814000041 ((01)00884838091474(21)US1814000041), US1814000043 ((01)00884838091474(21)US1814000043), US1814000044 ((01)00884838091474(21)US1814000044), US1814000045 ((01)00884838091474(21)US1814000045), US1814000046 ((01)00884838091474(21)US1814000046), US1814000047 ((01)00884838091474(21)US1814000047), US1814000048 ((01)00884838091474(21)US1814000048), US1814000049 ((01)00884838091474(21)US1814000049), US1814000050 ((01)00884838091474(21)US1814000050), US1814000051 ((01)00884838091474(21)US1814000051), US1814000052 ((01)00884838091474(21)US1814000052), US1814000053 ((01)00884838091474(21)US1814000053), US1814000054 ((01)00884838091474(21)US1814000054), US1814000055 ((01)00884838091474(21)US1814000055), US1814000056 ((01)00884838091474(21)US1814000056), US1814000058 ((01)00884838091474(21)US1814000058), US1814000059 ((01)00884838091474(21)US1814000059), US1814000060 ((01)00884838091474(21)US1814000060), US1814000061 ((01)00884838091474(21)US1814000061), US1814000062 ((01)00884838091474(21)US1814000062), US1814000063 ((01)00884838091474(21)US1814000063), US1814000064 ((01)00884838091474(21)US1814000064), US1814000065 ((01)00884838091474(21)US1814000065), US1814000066 ((01)00884838091474(21)US1814000066), US1814000067 ((01)00884838091474(21)US1814000067), US1814000068 ((01)00884838091474(21)US1814000068), US1814000069 ((01)00884838091474(21)US1814000069), US1814000070 ((01)00884838091474(21)US1814000070), US1814000072 ((01)00884838091474(21)US1814000072), US1814000074 ((01)00884838091474(21)US1814000074), US1814000078 ((01)00884838091474(21)US1814000078), US1814000081 ((01)00884838091474(21)US1814000081), US1814000082 ((01)00884838091474(21)US1814000082), US1814000083 ((01)00884838091474(21)US1814000083), US1814000084 ((01)00884838091474(21)US1814000084), US1814000085 ((01)00884838091474(21)US1814000085), US1814000086 ((01)00884838091474(21)US1814000086), US1814000087 ((01)00884838091474(21)US1814000087), US1814000088 ((01)00884838091474(21)US1814000088), US1814000089 ((01)00884838091474(21)US1814000089), US1814000092 ((01)00884838091474(21)US1814000092), US1814000094 ((01)00884838091474(21)US1814000094), US1814000096 ((01)00884838091474(21)US1814000096).
- 510(k) numbers
- ["K101571"]
- PMA numbers
- not on file
- Product code
- DQK
- Firm FEI
- 1051786
- Firm city
- Orlando
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28