Skip to the content

FDA Medical Device Recalls (openFDA)

AMO Manufacturing USA, LLC: iDesign, Model - G300: System AWS (International), System AWS (China), Advanced WaveScan Studio (United States); iDesign Refractive Studio Aberrometer (International) and (United States), Model: G301

Recall number
Z-0443-2022
Recalling firm
AMO Manufacturing USA, LLC
Product
iDesign, Model - G300: System AWS (International), System AWS (China), Advanced WaveScan Studio (United States); iDesign Refractive Studio Aberrometer (International) and (United States), Model: G301
Status
Open, Classified
Initiated
2021-10-21
Posted
not on file
Terminated
not on file
Reason
Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.
Root cause
Software design
Action
On 09/02/21, Microsoft "PrintNightmare" Vulnerability notices were sent to customers informing them of the upcoming software patches to remediate the vulnerability. This notice was posted on the recalling firm's website in the Security Advisories section. On 10/21/21 correction notices started being sent to customers. Customers were informed that Technical Service Representatives would contact then to schedule installation of the Windows OS patch. Customers were asked to do/ensure the following: 1) Complete the customer reply form, 2) To safeguard patients form potential harm and to inform them what to do in the event of product malfunction, customers were directed to specific parts of the Instructions for Use. 3) Confirm all measurements against baseline pre-operative manifest refraction and do not proceed when outside of recommended limits of difference, for sphere, cylinder and axis, as specified in the operators manual and training material. 4) Maintain manual records of patient measurements. 5) Ensure you comply with obligations under applicable privacy laws relating to putting in place safeguards in connection with personal information for patients. Customers with additional questions were encouraged to call Technical Support at 1 (800)- 511-0911.
Quantity
711
Distribution
US: CA, TX, FL, LA, MD, IA, MA, VA, SD, MN, MO, NC, TN, SC, WA, NV, OH, KY, IN, CO, NJ, OK, NH, UT, AR, NY, AK, ND, NE, IL, AL, HI, CT, DE, KS, GA, OR, MS, PA, MT, ME, MI, AZ, ID, WI, VT, NM, CA. OUS: Canada, Mexico, Russian Federation, Turkey, Brazil, Egypt, Iraq, Saudi Arabia, Jordan, Norway, Colombia, Germany, Malaysia, India, Oman, Azerbaijan, Canada, Slovakia, Lebanon, Spain, Ireland, Italy, South Korea, Bangladesh, Chile, Taiwan, Tunisia, Kazakhstan, Libya, Argentina, Singapore, Austria, Australia, Vietnam, Greece, Poland, Indonesia, Yemen, Great Britain, South Africa, Croatia, France, Czech Republic, Algeria, Thailand, Bulgaria, United Arab Emirates, Belgium, Japan, China, Nepal, Israel, Switzerland, Dominican Republic, Hong Kong, Portugal, Ecuador, Kuwait, Netherlands, Slovenia, Philippines
Lot, serial or model codes
G300 with Windows XP. G301 with Windows 7 Embedded System.
510(k) numbers
not on file
PMA numbers
["P930016"]
Product code
LZS
Firm FEI
3006695864
Firm city
Milpitas
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register