Elekta, Inc.: Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Version: n/a Product Description: The Intended Purpose of Leksell Vantage" Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.
- Recall number
- Z-0433-2026
- Recalling firm
- Elekta, Inc.
- Product
- Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Version: n/a Product Description: The Intended Purpose of Leksell Vantage" Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.
- Status
- Open, Classified
- Initiated
- 2025-09-30
- Posted
- 2025-11-05
- Terminated
- not on file
- Reason
- Due to increased risk that paint flakes from the engravings may separate from the device and there is the possibility that the flake may inadvertently enter the wound
- Root cause
- Process change control
- Action
- On 09/30/2025, the firm sent via email an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that specified serial numbers, there is an increased risk that paint flakes from the engravings may separate from the Leksell Vantage Arc System components. Separation of paint flakes is likely to occur during cleaning and sterilization. If paint flakes are not washed away during the cleaning and sterilization process, these could remain on the Leksell Vantage Arc System when used in surgery. This poses a risk of paint flakes entering the surgical area. This situation has not been observed in clinical practice, but the risk should be fully considered and mitigated.
- Quantity
- 338
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of AR, CA, CO, FL, ID, IL, LA, MA, MI, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, WA, WI, and WV. The countries of Algeria, Australia, Austria, Belgium, China, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Republic of Korea, Kuwait, Lithuania, Mexico, Netherlands, New Zealand, Norway, Pakistan, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
- Lot, serial or model codes
- Lot Code: UDI References: Arc: 07340048304856 Support Arc: 07340048304849 Arcs and Support Arcs with Serial Numbers: SH00001-SH00338
- 510(k) numbers
- ["K190887"]
- PMA numbers
- not on file
- Product code
- HAW
- Firm FEI
- 1037831
- Firm city
- Atlanta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28