Siemens Healthcare Diagnostics, Inc.: Siemens Dimension HB1C Flex reagent cartridge; Dimension Hemoglobin A1c (HB1C) Flex reagent cartridge on the Dimension clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long-term glucose control in individuals with diabetes mellitus.

Recall number
Z-0420-2015
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Siemens Dimension HB1C Flex reagent cartridge; Dimension Hemoglobin A1c (HB1C) Flex reagent cartridge on the Dimension clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long-term glucose control in individuals with diabetes mellitus.
Status
Terminated
Initiated
2014-09-30
Posted
2014-11-21
Terminated
2015-08-10
Reason
Certain lots of the reagent cartridges show confirmed positive bias averaging 0.4% Hemoglobin A1c units and occasionally up to 1.0% HbA1c units for patient samples when compared to the National Glycohemoglobin Standardization Program (NGSP). QC samples may exhibit a similar bias.
Root cause
Nonconforming Material/Component
Action
Siemens sent an Urgent Medical Device Correction letter dated September 2014 was provided to all Dimension Vista HbA1c and Dimension HB1C customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers. Customers were instructed to discontinue use of and discard the affected devices. Customers need to review the letter with their Medical Director. Siemens will replace any unused inventoryof the affected lots at no charge. Customers were asked to complete the attached form and fax it to 302-631-7795. Customers should retain a copy of the letter with their laboratory records, and forward the letter to those who may have received this product. Customers with questions should contact their Siemens Customer Care Center - Technical Solutions or their local Siemens technical support representative. For questions regarding this recall call 800-441-9250.
Quantity
53, 046
Distribution
Worldwide Distribution: US (nationwide) and internationally to: Canada.
Lot, serial or model codes
Siemens Dimension Vista HbA1c Flex reagent cartridge: Catalog Number K3105A, Siemens Material Number (SMN) 10470481, Lot Numbers 13294BA 13308BA 13329AA 13350AA 14006AA 14041BA 14069BB 14083BA   Siemens Dimension HB1C Flex reagent cartridge: Catalog Number DF105A, Siemens Material Number (SMN) 10483822, Lot Numbers GA4266 BA4273 BA4280 BA4287 GA4301 GA4315 GC4322 GA4343 GA4350 GA4357 GA5013 GA5020
510(k) numbers
K102045, K102510
PMA numbers
not on file
Product code
LCP
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source