Baxter Healthcare Corp.: Baxter 6060 Multi-Therapy Infusion Pump, product codes 2M9832, 2M9832P and 2M9832R, and Sabratek 6060 Homerun Infusion Pumps, product codes 606000, 606000-40, 606000-40L, and 606000-40I; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore

Recall number
Z-0412-06
Recalling firm
Baxter Healthcare Corp.
Product
Baxter 6060 Multi-Therapy Infusion Pump, product codes 2M9832, 2M9832P and 2M9832R, and Sabratek 6060 Homerun Infusion Pumps, product codes 606000, 606000-40, 606000-40L, and 606000-40I; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore
Status
Terminated
Initiated
2005-11-14
Posted
2006-01-28
Terminated
2007-06-19
Reason
Failures within the PCA profile as well as incidents resulting in interruptions of therapy in various profiles.
Root cause
Other
Action
Baxter sent recall letters dated 11/14/05 to all of their 6060 pump customers on the same date via DHL overnight delivery. The accounts were informed of the reports of failures within the PCA profile and interruptions of therapy in various profiles. Due to the obsolescence of certain critical components, Baxter is conducting a controlled removal of all 6060 pumps from the market. Baxter will coordinate with customers individually to ensure a smooth transition to a substitute device. Baxter provided interim instructions to discontinue use of the PCA profile to avoid over infusion or non-delivery conditions, and to discontinue the use of the pump for therapies where interruption of the infusion could cause immediate patient harm. Baxter representatives telephoned each account within 48 hours of the letter, and visited them within a week to help the customers transition to other infusion pumps. A press release issued on 11/15/05. Baxter sent a follow-up recall letter to the accounts on 6/20/06, informing them that service repairs for the 6060 Infusion Pump will cease after 12/31/06, and compensation requests cannot be processed for pumps received after 12/31/06. Production of sets for the 6060 will cease on 10/1/06, and will not be available after 12/31/06. The accounts were requested to complete and fax back to Baxter the enclosed reply form, indicating the current inventory of 6060 pumps remaining at the facility.
Quantity
34,000 pumps
Distribution
Nationwide and internationally to Belgium, Germany, Finland, France, the United Kingdom, Ireland, Sweden, Norway, Canada, Japan, New Zealand and Australia.
Lot, serial or model codes
All serial numbers of all product codes
510(k) numbers
["K002679"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register