Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Molecular, Inc.: Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System to detect and quantitate hepatitis C virus (HCV) RNA in human serum or plasma.

Recall number
Z-0407-2024
Recalling firm
Abbott Molecular, Inc.
Product
Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System to detect and quantitate hepatitis C virus (HCV) RNA in human serum or plasma.
Status
Open, Classified
Initiated
2023-10-30
Posted
2023-11-29
Terminated
not on file
Reason
Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.
Root cause
Other
Action
Abbott issued an Urgent Field Safety Notice to its consignees on 10/30/2023 via letter delivered using FedEx priority overnight express. The notice explained the problem, the potential impact/risk to patient, and requested the following actions: -discard all affected product -notify all those to which the products were transferred -follow internal laboratory procedures in the case of suspected incorrect results Credit or replacement will be issued for units discarded/destroyed. Abbott expanded the recall and notified consignees on 07/23/2024 via letter. The letter instructed consignees of an additional lot number included in the recall, but otherwise contained the same instructions.
Quantity
1,881 units
Distribution
Worldwide
Lot, serial or model codes
a) 08N50-095: UDI/DI 00884999047648, Lot Numbers: 381582, 382448, 383204; b) 08N50-090: UDI/DI 00884999047761, Lot Numbers: 383203, 385720, 381581, 382447
510(k) numbers
not on file
PMA numbers
["P190025"]
Product code
MZP
Firm FEI
3005248192
Firm city
Des Plaines
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register