Skip to the content

FDA Medical Device Recalls (openFDA)

Sonosite, Inc.: SonoSite L25 Needle Guide Kit, 21 GAUGE, Part Number P04053-01. Manufactured for: SonoSite, Inc., Bothell, WA 98021 Each kit is packaged in a hermetically sealed clear plastic pouch with white backing. Product is labeled as Sterile. 24 pouches are packaged in a white cardboard box and labeled as SonoSite L25 Needle Guide Kit, 21 GAUGE REF# P04053-01, Quantity: 24, CONTENTS STERILE IF PACKAGE IS INTACT. Mfg for: Sonosite, Inc. 888/482-9449*425/951-1200, Bothell, WA 98021 USA. www.sonosite.com

Recall number
Z-0407-06
Recalling firm
Sonosite, Inc.
Product
SonoSite L25 Needle Guide Kit, 21 GAUGE, Part Number P04053-01. Manufactured for: SonoSite, Inc., Bothell, WA 98021 Each kit is packaged in a hermetically sealed clear plastic pouch with white backing. Product is labeled as Sterile. 24 pouches are packaged in a white cardboard box and labeled as SonoSite L25 Needle Guide Kit, 21 GAUGE REF# P04053-01, Quantity: 24, CONTENTS STERILE IF PACKAGE IS INTACT. Mfg for: Sonosite, Inc. 888/482-9449*425/951-1200, Bothell, WA 98021 USA. www.sonosite.com
Status
Terminated
Initiated
2005-12-12
Posted
2006-01-19
Terminated
2006-07-31
Reason
22 gauge needle guides were intermixed with 21 gauge guides and labeled as 21 gauge needle guides.
Root cause
Other
Action
Affected customers were contacted via telephone between December 12-19, 2005. Provided instructions for returning defective product. Notification letter was distributed to all customers on December 14, 2005 instructing customers to discontinue use of the needle guide and return remaining portion.
Quantity
355 boxes/24 kits per box
Distribution
Product was distributed to distributors and hospitals nationwide and internationally.
Lot, serial or model codes
Box Lot Code: M778940; M779110; M785430; M790190; M791960; M793010; M793020; M793530; M799610; M799620; M803570; M803580; M806400; M806410; M806420; M809020; M810990; M823040; M830430.  Pouch Lot Code: M761640; M778400; M781250; M783140; M785270; M791720; M797960; M801010; M802700; M804630; M807010; M811660; M831060; M832790
510(k) numbers
["K030064"]
PMA numbers
not on file
Product code
ITX
Firm FEI
3002087856
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register