Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes USA HQ, Inc.: Synthes Material Mix (Adjustable Cervical Distractor-Right, Matrix Distractor Rack, T-PAL Spacer Remover, Implant Inserter) Synthes Material Mix: Adjustable Cervical Distractor-Right is used to distract spine segment. Matrix Distractor Rack is used to distract the interbody disc space. T-PAL Spacer Remover is used to remove T-PAL spacer. Implant Inserter is used to insert the ProDisc implant.

Recall number
Z-0393-2014
Recalling firm
Synthes USA HQ, Inc.
Product
Synthes Material Mix (Adjustable Cervical Distractor-Right, Matrix Distractor Rack, T-PAL Spacer Remover, Implant Inserter) Synthes Material Mix: Adjustable Cervical Distractor-Right is used to distract spine segment. Matrix Distractor Rack is used to distract the interbody disc space. T-PAL Spacer Remover is used to remove T-PAL spacer. Implant Inserter is used to insert the ProDisc implant.
Status
Terminated
Initiated
2013-10-15
Posted
2013-11-26
Terminated
2015-08-31
Reason
One material batch used for production of the affected parts of the Synthes Material Mix were mixed in with an incorrect material with a lower than specified hardness and corrosion resistance.
Root cause
Nonconforming Material/Component
Action
SYNTHES sent an Urgent Notice: Medical Device Recall letter dated October 15, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to call Synthes at 1-800-479-6329 to obtain a Return Authorization Number. Complete the Verification Section at the end of the letter by checking the appropriate box indicating affected product was located. Customers were also asked to indicate the number of devices found and note the Return Authorization Number. Return the Verification Form with the product to: Credit/Returns Synthes 1101 Synthes Avenue Monument, CO 80132 Customers should also return the Verification Form if no affected product was found in their inventory. Return all documents to Synthes by fax: 888-679-4875 or Scan/email: FieldAction@synthes.com Customers with questions were instructed to call 610-719-5450 or contact their Synthes Sales Consultant. For questions regarding this recall call 610-719-5000.
Quantity
approximately 44
Distribution
Worldwided Distribution: USA (nationwide) including IL, MO, TX, WA and Internationally to Canada and Switzerland.
Lot, serial or model codes
part no. 396.396 (lot nos. T989765 and T989780), part no. 03.632.087 (lotT987270), part no. 03.812.005 (lot nos. T987734, T988931, and T987623), and part no. 03.820.129 (lot no. T988206).
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3005180112
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register