Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Logix-CM Compounder Software, catalog 2M8400; Baxter Healthcare Corporation, Medication Delivery System, Round Lake, IL 60073

Recall number
Z-0392-05
Recalling firm
Baxter Healthcare Corp.
Product
Logix-CM Compounder Software, catalog 2M8400; Baxter Healthcare Corporation, Medication Delivery System, Round Lake, IL 60073
Status
Terminated
Initiated
2004-12-08
Posted
2004-12-30
Terminated
2005-11-01
Reason
A software anomaly in the Logix CM software could result in a drug mixture being compounded without one of the ingredients if two drugs have the same indentifier in Logix and another order entry program being used with the Logix software.
Root cause
Other
Action
Urgent Device Correction letters dated 12/8/04 were sent to the end users via overnight mail. The accounts were informed of the potential for omission of an ingredient if there are duplicate drug Id''s in the order, and were requested to complete and fax the customer reply form to Baxter at 847-270-5457, indicating if they are currently using Logix-CM software. Upon receipt of the reply form, Baxter will schedule a visit to the facility to examine the base and additive inventory information in their MOS database to confirm that there are no duplicate Drug ID numbers and that this issue has not occurred at the facility. They will also secure the MOS database to ensure that the situation will not be able to occur. Any questions were directed to Baxter Clintec Nutrition, option 1, at 1-800-422-2751.
Quantity
12 units
Distribution
Maryland, Arizona, Indiana, Ohio, Minnesota, Pennsylvania and internationally to Canada
Lot, serial or model codes
Software version 2.0.9 for Clintec Compounders
510(k) numbers
not on file
PMA numbers
not on file
Product code
LHI
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register