Skip to the content

FDA Medical Device Recalls (openFDA)

Omron Healthcare, Inc.: Medication Bottle (U22-3), a component of the MicroAir Vibrating Mesh Nebulizer, Model NE-U22V/NE-22VAC; Distributed by: Omron Healthcare, Inc., 300 Lakeview Parkway, Vernon Hills, IL 60061

Recall number
Z-0382-05
Recalling firm
Omron Healthcare, Inc.
Product
Medication Bottle (U22-3), a component of the MicroAir Vibrating Mesh Nebulizer, Model NE-U22V/NE-22VAC; Distributed by: Omron Healthcare, Inc., 300 Lakeview Parkway, Vernon Hills, IL 60061
Status
Terminated
Initiated
2004-11-02
Posted
2005-01-05
Terminated
2007-06-26
Reason
A potential failure of the vibrating mechanism of the medication bottle which causes a failure to transmit vibration and results in no nebulization.
Root cause
Other
Action
Omron sent 'Removal Notices' dated 11/2/04 to all of their direct accounts receiving the nebulizers during the affected time frame. The accounts were informed of the potential failure of the vibrating mechanism, and listed the serial numbers of the nebulizers that may contain the affected medication bottles. Medication bottles with either a blue dot near the bottle lever or a blue O-ring around the body of the bottle were manufactured after corrections were made and are not subject to the recall. Affected product should be returned to Omron for replacement. Any questions were directed to Customer Service at 800-231-4030.
Quantity
63,560 units
Distribution
Nationwide
Lot, serial or model codes
Bottle lots 43A, 53A, 63A, 73A, 83A, 93A, X3A, Y3A and Z3A, a component of Nebulizer model NE-U22 serial numbers 3400001A and higher, 3500001A and higher, 3600001A and higher, 3700001A and higher, 3800001A and higher, 3900001A and higher, 3X00001A and higher, 3Y00001A and higher, and 3Z00001A through 3Z00696A. Medication bottles with either a blue dot or a blue O-ring were manufactured after corrections were made and are not subject to the recall.
510(k) numbers
["K923024"]
PMA numbers
not on file
Product code
CAF
Firm FEI
1450057
Firm city
Bannockburn
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register