Skip to the content

FDA Medical Device Recalls (openFDA)

Canon Medical System, USA, INC.: Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.

Recall number
Z-0360-2020
Recalling firm
Canon Medical System, USA, INC.
Product
Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.
Status
Terminated
Initiated
2019-02-08
Posted
not on file
Terminated
2022-06-07
Reason
Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.
Root cause
Error in labeling
Action
On 02/08/2019, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification to customers via USPS return receipt mail informing them of the potential problem with the Catheterization Table (CAT-870B) may tilt without operator involvement due to accidental liquid spillage. Customers are instructed to: a. Please keep the tableside console and the x-ray detector of the support units covered with sterile caps as stated in the Caution section of the operation manual. The caps are there to prevent liquids from coming into contact with the tableside console and the x-ray detector. b. If an actual event occurs, please push the emergency switch and call your service representative. c. Please share this information with all users and reviewing radiologist as well as clinical engineering and the biomedical group at your facility. d. Please complete and return the attached form and fax it to the toll free number 877-349-3054. This form may also be sent via email to RAffairs@us.medical.canon. If customers have any questions regarding this letter, they are free to contact Regulatory Affairs Manager at (800) 421-1968 or their local Representative at (800) 521-1968.
Quantity
9 systems
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of Netherlands.
Lot, serial or model codes
Model CAT-870B D053991  Serial No. Manufacture Date 99D08X2101 October 2008 99D0582039 August 2005 99D0762072 June 2007 99D0852090 May 2008 99D07Y2078 November 2007 99D0812081 January 2008 99D0582042 August 2005 99D0552033 May 2005 D551206 May 2005
510(k) numbers
["K081624"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.