Skip to the content

FDA Medical Device Recalls (openFDA)

Argon Medical Devices, Inc: SuperCore Biopsy Instrument: REF/Product Description: 701114090/SUPERCORE BIOPSY INSTR 14GA X 9CM; 701114150/SUPERCORE BIOPSY INSTR 14GA X 15CM; 701116090/SUPERCORE BIOPSY INSTR 16GA X 9CM; 701118060/SUPERCORE BIOPSY INSTR 18GA X 6CM; 701118090/SUPERCORE BIOPSY INSTR 18GA X 9CM; 701118150/SUPERCORE BIOPSY INSTR 18GA X 15CM; 701118200/SUPERCORE BIOPSY INSTR 18GA X 20CM; 701120090/SUPERCORE BIOPSY INSTR 20GA X 9CM; 701120150/SUPERCORE BIOPSY INSTR 20GA X 15CM; 701120200/SUPERCORE BIOPSY INSTR 20GA X 20CM; 701214090/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 9CM; 701214150/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 15CM; 701216090/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 9CM; 701216150/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 15CM; 701218090/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 9CM; 701218150/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 15CM; 701218200/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 20CM; 701220090/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 9CM; 701220150/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 15CM; 701220200/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 20CM; 701114090M/SUPERCORE BIOPSY INSTR 14GA X 9CM-MAMMOTOME; 701114150M/SUPERCORE BIOPSY INSTR 14GA X 15CM-MAMMOTOME; 701116090M/SUPERCORE BIOPSY INSTR 16GA X 9CM-MAMMOTOME; SC 1409/PGI Supercore 14GA x 9CM; SC 1609/PGI Supercore 16GA x 9CM; SC 1806/PGI Supercore 18GA x 6CM;

Recall number
Z-0354-2022
Recalling firm
Argon Medical Devices, Inc
Product
SuperCore Biopsy Instrument: REF/Product Description: 701114090/SUPERCORE BIOPSY INSTR 14GA X 9CM; 701114150/SUPERCORE BIOPSY INSTR 14GA X 15CM; 701116090/SUPERCORE BIOPSY INSTR 16GA X 9CM; 701118060/SUPERCORE BIOPSY INSTR 18GA X 6CM; 701118090/SUPERCORE BIOPSY INSTR 18GA X 9CM; 701118150/SUPERCORE BIOPSY INSTR 18GA X 15CM; 701118200/SUPERCORE BIOPSY INSTR 18GA X 20CM; 701120090/SUPERCORE BIOPSY INSTR 20GA X 9CM; 701120150/SUPERCORE BIOPSY INSTR 20GA X 15CM; 701120200/SUPERCORE BIOPSY INSTR 20GA X 20CM; 701214090/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 9CM; 701214150/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 15CM; 701216090/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 9CM; 701216150/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 15CM; 701218090/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 9CM; 701218150/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 15CM; 701218200/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 20CM; 701220090/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 9CM; 701220150/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 15CM; 701220200/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 20CM; 701114090M/SUPERCORE BIOPSY INSTR 14GA X 9CM-MAMMOTOME; 701114150M/SUPERCORE BIOPSY INSTR 14GA X 15CM-MAMMOTOME; 701116090M/SUPERCORE BIOPSY INSTR 16GA X 9CM-MAMMOTOME; SC 1409/PGI Supercore 14GA x 9CM; SC 1609/PGI Supercore 16GA x 9CM; SC 1806/PGI Supercore 18GA x 6CM;
Status
Open, Classified
Initiated
2021-10-04
Posted
not on file
Terminated
not on file
Reason
Due to component (housing and plunger) detaching from the semi-automatic biopsy instruments.
Root cause
Device Design
Action
On 09/30/2021, the firm sent an "Urgent-Product Recall Notice" letter to customers via US Postal Service Registered Mail informing them that the firm has received complaints regarding the semi-automatic Biopsy instrument coming apart during shipping or prior to use. According to the firm's investigation, the plastic housing and plunger can be separated more easily than normal for the lots manufactured during this time frame. Customers are instructed to return all affected product to the Recalling Firm's Athens TX facility using RGA#26683, attention Andrea Wieczor. Mailing address is: RGA# 26683 Argon Medical Devices, Inc. 1445 Flat Creek Road Athens, TX 75751 USA Att: Andrea Wieczor In addition, Customers are instructed to complete the response acknowledgement form included with the letter/email and return by fax, mail, or email as soon as possible. For questions. contact the Director, Post-Market Experience at email address Brian.Rogers@argonmedical.com, or contact Arbee Cummings at Arbee.Cummings@argonmedical.com or Andrea Wieczor at Andrea.Wieczor@argonmedical.com. (Updated information 11/2/2021), on 10/28/2021, the firm sent an updated "Urgent-Product Recall Notice for Additional Lots" letter via mail informing customers that they are expanding the recall to additional lots (list on the Inventory Sheet attached to recall notice). Again the Recalling Firm is instructing customer to return the affected product (listed on the Inventory Sheet).
Quantity
79,910 devices
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. The countries of ALB, ARE, ARG, AUS, AUT, BEL, BGD, BRA, CAN, CHE, CHL, CHN, COL, CZE, DEU, ECU, ESP, EST, FRA, GBR, GRC, HKG, HUN, IND, IRN, ISR, ITA, JOR, JPN, MAR, MEX, MYS, NLD, NPL, NZL, PAK, PER, PHL, POL, PRT, SAU, SGP, SRB, THA, TUR, and TWN.
Lot, serial or model codes
REF/Product Description: 701114090/SUPERCORE BIOPSY INSTR 14GA X 9CM; Lot Numbers: 11364350 11365643 11368388 11371872 11374998 11376772 11378864 11382837 11385962 11389848 701114150/SUPERCORE BIOPSY INSTR 14GA X 15CM; Lot Numbers: 11368389; 11376773; 701116090/SUPERCORE BIOPSY INSTR 16GA X 9CM; Lot Numbers: 11362716; 11363320; 11365935; 11377634; 11381340; 11384192; 11388461; 11388552; 11389139; 701118060/SUPERCORE BIOPSY INSTR 18GA X 6CM; Lot Numbers: 11361702; 11365788; 11371012; 11374235; 11376598; 11379722; 11382768; 11383890; 11388656; 11389138; 701118090/SUPERCORE BIOPSY INSTR 18GA X 9CM; Lot Numbers: 11362759; 11363229; 11364420; 11367439; 11370106; 11372917; 11376074; 11377430; 11378217; 11383193; 11383785; 11384243; 11387238; 11388864; 701118150/SUPERCORE BIOPSY INSTR 18GA X 15CM; Lot Numbers: 11370107; 11375663; 11377819; 11382602; 11385139; 11385965; 11389140; 11390503; 701118200/SUPERCORE BIOPSY INSTR 18GA X 20CM; Lot Numbers: 11372225; 11374410; 11377431; 11384193; 11388505; 701120090/SUPERCORE BIOPSY INSTR 20GA X 9CM; Lot Numbers: 11374411; 11378865; 11383280; 701120150/SUPERCORE BIOPSY INSTR 20GA X 15CM; Lot Numbers: 11366252 11367440; 11376374; 11377820; 11381387; 11385960; 701120200/SUPERCORE BIOPSY INSTR 20GA X 20CM; Lot Numbers: 11366253; 11380018; 11388893; 701214090/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 9CM; Lot Numbers: 11369916; 11378293; 11383587; 11387065; 701214150/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 15CM; Lot Numbers: 11377967; 11382843; 701216090/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 9CM; Lot Numbers: 11364421; 11367443; 11374684; 11379724; 11385966; 11388551; 701216150/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 15CM; Lot Numbers: 11382845; 11388550; 11390251; 701218090/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 9CM; Lot Numbers: 11361701; 11364247; 11367445; 11372366; 11374807; 11378687; 11383192; 11383784; 11388286; 701218150/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 15CM; Lot Numbers: 11371404; 11373084; 11377636; 11380993; 11383194; 11383195; 11387400; 11387833; 11390504; 701218200/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 20CM; Lot Numbers: 11366315; 11372227; 11377432; 11381193; 11382998; 11384716; 701220090/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 9CM; Lot Numbers: 11367446; 11370574; 11375664; 11380589; 11381572; 11388507; 701220150/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 15CM; Lot Numbers: 11368750; 11371876; 11372228; 11376375; 11377429; 11380104; 11384717; 11387148; 701220200/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 20CM; Lot Numbers: 11362655; 11363516; 11372322; 11378295; 11380103; 11382884; 11387584; 701114090M/SUPERCORE BIOPSY INSTR 14GA X 9CM-MAMMOTOME; Lot Numbers: 11371356; 11382375; 11389615; 701114150M/SUPERCORE BIOPSY INSTR 14GA X 15CM-MAMMOTOME; Lot Number: 11371357; 701116090M/SUPERCORE BIOPSY INSTR 16GA X 9CM-MAMMOTOME; Lot Number: 11371359; SC 1409/PGI Supercore 14GA x 9CM; Lot Numbers: 11370550; 11384881; SC 1609/PGI Supercore 16GA x 9CM; Lot Number: 11370552; SC 1806/PGI Supercore 18GA x 6CM; Lot Numbers: 11375600; 11364469;
510(k) numbers
["K974814"]
PMA numbers
not on file
Product code
KNW
Firm FEI
1625425
Firm city
Athens
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register