Skip to the content

FDA Medical Device Recalls (openFDA)

Cytocell Ltd.: CytoCell BCL11B Proximal in FITC Spectrum-LDT for investigation T-cell related malignancy. This device is an analyte specific reagent probe covering the 14q32.2 locus proximal to the BCL11B gene on chromosome 14. Ref: MPD39781

Recall number
Z-0353-2023
Recalling firm
Cytocell Ltd.
Product
CytoCell BCL11B Proximal in FITC Spectrum-LDT for investigation T-cell related malignancy. This device is an analyte specific reagent probe covering the 14q32.2 locus proximal to the BCL11B gene on chromosome 14. Ref: MPD39781
Status
Terminated
Initiated
2022-10-27
Posted
2022-12-01
Terminated
2024-12-27
Reason
Device is mislabeled. The package insert for this ASR device contains a chromomap - an image showing the position the FISH probe maps to on the specific chromosome. For this device the chromomap in version 1 of the package insert is incorrect and doesn't not match the probe design or specification contained on the package insert
Root cause
Under Investigation by firm
Action
Cytocell issued Urgent Medical Device Recall letter on 10/27/22 to the end-user via email. Letter states reason for recall, health risk and action to take: Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that all users review and implement the enclosed new package insert (v2) for this device. Please dispose of any copies of v1 of the package insert. Once implemented, all devices can continue to be used as the device itself has not malfunctioned. Please complete and return the enclosed response form as soon as possible. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com.
Quantity
not on file
Distribution
TX
Lot, serial or model codes
Lot RD22/230/10
510(k) numbers
not on file
PMA numbers
not on file
Product code
MVU
Firm FEI
3002646096
Firm city
Cambridge
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register