Skip to the content

FDA Medical Device Recalls (openFDA)

Urologix, Inc.: Prostatron Praktis Type II for Transurethral Microwave Thermotherapy. The Prostatron is a computer-controlled device designed to deliver microwave energy to the prostrate for the treatment of BPH (Benign Prostatic Hyperplasia). The Prostatron utilizes a transurethral microwave antenna, with simultaneous urethral cooling, to heat the prostate. This heating process is regulated through feedback from a series of temperature sensors.

Recall number
Z-0345-04
Recalling firm
Urologix, Inc.
Product
Prostatron Praktis Type II for Transurethral Microwave Thermotherapy. The Prostatron is a computer-controlled device designed to deliver microwave energy to the prostrate for the treatment of BPH (Benign Prostatic Hyperplasia). The Prostatron utilizes a transurethral microwave antenna, with simultaneous urethral cooling, to heat the prostate. This heating process is regulated through feedback from a series of temperature sensors.
Status
Terminated
Initiated
2003-12-10
Posted
2004-07-20
Terminated
2005-11-15
Reason
During treatments, using the Prostatron units, the devices' Temperature Sensing System may change the rectal temperature reading by up to 2 degrees Celsius and may briefly delay updates to the rectal temperature reading.
Root cause
Other
Action
A letter about the problem, actions to take to avoid the problem, and the proposed software correction was sent to consignees on December 10, 2003. A software upgrade will be performed on all units in March 2004.
Quantity
146 units
Distribution
The product was shipped to consignees located nationwide in the United States and to foreign consignees in Australia, France and Japan.
Lot, serial or model codes
Serial numbers PP80023 through PP80040 and PP037 through PP180
510(k) numbers
not on file
PMA numbers
["P950014"]
Product code
MEQ
Firm FEI
3000718675
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register