Skip to the content

FDA Medical Device Recalls (openFDA)

SCC Soft Computer: SoftLab/Mic ASCII Add-on Versions 4.0.1.14, 4.0.1.15, and 4.0.1.16. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Version 4.0.1.14 Manufactured & Distributed 12/29/2006. Version 4.0.1.15 Manufactured & Distributed 05/02/2007. Version 4.0.1.16 Manufactured & Distributed 08/22/2008. Used in a medical research or clinical laboratory setting by knowledgeable, trained and experienced personnel.

Recall number
Z-0344-2011
Recalling firm
SCC Soft Computer
Product
SoftLab/Mic ASCII Add-on Versions 4.0.1.14, 4.0.1.15, and 4.0.1.16. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Version 4.0.1.14 Manufactured & Distributed 12/29/2006. Version 4.0.1.15 Manufactured & Distributed 05/02/2007. Version 4.0.1.16 Manufactured & Distributed 08/22/2008. Used in a medical research or clinical laboratory setting by knowledgeable, trained and experienced personnel.
Status
Terminated
Initiated
2010-02-16
Posted
2010-11-10
Terminated
2012-07-17
Reason
When a test is being verified, the wrong (expired) ranges may be associated to that test result, causing an incorrect flag to be posted for that test result.
Root cause
Software design
Action
Clients were notified via the proprietary communication software on 02/16/2010. The Notification Verbiage identified the affected product and informed clients of the potential hazard, root cause, corrective action, and requested action. Clients were instructed to acknowledge receipt of the notification. If the client wishes to receive the correction, they should document a request to load the software.
Quantity
31
Distribution
Nationwide Distribution - US and Puerto Rico.
Lot, serial or model codes
Versions 4.0.1.14, 4.0.1.15, and 4.0.1.16.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1000307651
Firm city
Clearwater
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source