Skip to the content

FDA Medical Device Recalls (openFDA)

WISSNER-BOSSERHOFF GMBH: SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198

Recall number
Z-0329-2025
Recalling firm
WISSNER-BOSSERHOFF GMBH
Product
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198
Status
Open, Classified
Initiated
2022-07-11
Posted
2024-11-08
Terminated
not on file
Reason
When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.
Root cause
Device Design
Action
Wissner-bosserhoff issued an Urgent safety information notice along with supplement use instructions notice to its sole US consignee on 07/11/2022. The notice explained the problem, risk to patient, and requested the user update the IFU for the device. Distributors were directed to notify their customers.
Quantity
77 units
Distribution
US Nationwide distribution in the states of NC.
Lot, serial or model codes
UDI/DI 4050621100401, 19-0990 UDI/DI 4050621100012, 19-1062 UDI/DI 4050621100432, 19-1152 UDI/DI 4050621100449, 19-1119 UDI/DI 4050621100470, 19-1155 UDI/DI 4050621100487, 19-1156
510(k) numbers
not on file
PMA numbers
not on file
Product code
LLI
Firm FEI
3015979371
Firm city
Wickede Ruhr
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register