Encore Medical, Lp: Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior stabilized implant.

Recall number
Z-0325-2017
Recalling firm
Encore Medical, Lp
Product
Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior stabilized implant.
Status
Terminated
Initiated
2016-09-19
Posted
not on file
Terminated
2017-10-20
Reason
It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery. The current design of the captured box cut guides can create an instance where the surgeon is not fully completing the box cut on the side.
Root cause
Under Investigation by firm
Action
The recalling firm sent a recall notification letter to affected consignees on 9/19/16. The field safety notice provided a newly revised Surgical Technique to educate users on proper use of the device.
Quantity
436 units
Distribution
US
Lot, serial or model codes
202351L01, 202434L01, 203779L01, 202351L02, 202434L02, 202351L03, 202434L03, 203779L03, 202351L04, 202434L04, 202351L05, 202434L05, 202351L06, 202434L06, 203779L06, 202351L07, 202434L07, 203779L07, 202351L08, 202434L08, 203779L08, 202351L09, 202434L09, 202351L10, 202434L10, 203203L01
510(k) numbers
["K160342"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register