Becton Dickinson Infusion Therapy Systems Inc.: BD Nexiva, 20 GA 1.00 IN (1.1 X 25 mm), 3660 ml/hr. 61ml/min; Closed IV Catheter System- Dual port -61 mL/min (3660 mL/hr)-0.5mL, REF 383536; BD Q-Style Luer Access Split Septum BD Vialon Material; BD Instaflash Needle Technology; Rx Only, Sterile EO. BD Nexiva closed IV catheter systems are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon" catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, dual port adapter, vent plug and pre-attached BD Q-Syte Luer Access Split Septum. The needle and catheter are protected by a needle cover. A BD Q-Syte device or an end cap with protective cover is provided in the unit package. The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury. These devices have BD Instaflash" Needle Technology, allowing for immediate visualization of blood along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing during insertion. Both the stabilization platform and dual port adapter are color coded to indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue, 20GA (1.1 mm)=Pink, 18GA (1.3 mm)=Green). BD Nexiva closed IV catheter systems are intended to be inserted into a patient s vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.

Recall number
Z-0306-2022
Recalling firm
Becton Dickinson Infusion Therapy Systems Inc.
Product
BD Nexiva, 20 GA 1.00 IN (1.1 X 25 mm), 3660 ml/hr. 61ml/min; Closed IV Catheter System- Dual port -61 mL/min (3660 mL/hr)-0.5mL, REF 383536; BD Q-Style Luer Access Split Septum BD Vialon Material; BD Instaflash Needle Technology; Rx Only, Sterile EO. BD Nexiva closed IV catheter systems are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon" catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, dual port adapter, vent plug and pre-attached BD Q-Syte Luer Access Split Septum. The needle and catheter are protected by a needle cover. A BD Q-Syte device or an end cap with protective cover is provided in the unit package. The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury. These devices have BD Instaflash" Needle Technology, allowing for immediate visualization of blood along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing during insertion. Both the stabilization platform and dual port adapter are color coded to indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue, 20GA (1.1 mm)=Pink, 18GA (1.3 mm)=Green). BD Nexiva closed IV catheter systems are intended to be inserted into a patient s vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.
Status
Open, Classified
Initiated
2021-09-08
Posted
not on file
Terminated
not on file
Reason
There is a breach in the product packaging that renders the product non-sterile.
Root cause
Packaging process control
Action
On September 30, 2021, the firm sent out a customer notification letter "Urgent Medical Device Recall" to their customers via Fed-Ex with the instruction due to the packaging breach, identified during the in house internal leak testing. This breach in the product packaging rendered the enclosed IV catheters non-sterile. Holes have been detected on the bottom tray of device unit packages. The notification stated to destroy the devices which are present in customers/distributors inventory. The recalling firm notified in a customer notification letter to fill and send the customer response form with customer's details and acknowledgement of destruction of the device list , lot number and quantities for the replacement. The contact information will be provided for the customers to call : 1-844-8BD-LIFE (1-844-823-5433)Say Recall when prompted, Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com The customer response form to be returned via: E-Mail BDRC17@bd.com, Fax No. is 312-949-0375.
Quantity
155,840 devices, 80 units in each box
Distribution
U.S. Nationwide Distribution: AR, AZ, MA, GA, NC, OH, WA, NY, NJ, HI, CA, MD, RI, MT, IN, PA, IL OK, OR, AK, CT, TX, ME, MO, FL, VA, LA, CO, UT O.U.S.: None
Lot, serial or model codes
Catalog number/Model number: 383536; Batch numbers/UDI numbers: Lot: 1193055/(17)240630(10)1193055(30)80/(01)50382903835367, 2024-06-30;  Lot: 1166273: (17)240630(10)1166273(30)80/(01)50382903835367, 2024-06-30 Lot: 1166273: (17)240630(10)1166273(30)80/(01)50382903835367, 2024-06-30 Lot: 1188953: (17)240630(10)1188953(30)80/(01)50382903835367, 2024-06-30
510(k) numbers
["K183399"]
PMA numbers
not on file
Product code
FOZ
Firm FEI
1710034
Firm city
Sandy
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register