Skip to the content

FDA Medical Device Recalls (openFDA)

Zest Anchors Inc: "LOCATOR Abutment for 5.7 Screw-Vent & Compatibles." Zest Anchors, Inc. Product is sold for use with any Zimmer Dental 5.7 mm diameter Tapered Screw-Vent Implant. The product is used with supra-gingival, universal hinge, resilient overdenture attachment for endosseous implants.

Recall number
Z-0298-2009
Recalling firm
Zest Anchors Inc
Product
"LOCATOR Abutment for 5.7 Screw-Vent & Compatibles." Zest Anchors, Inc. Product is sold for use with any Zimmer Dental 5.7 mm diameter Tapered Screw-Vent Implant. The product is used with supra-gingival, universal hinge, resilient overdenture attachment for endosseous implants.
Status
Terminated
Initiated
2008-04-09
Posted
2008-12-01
Terminated
2008-12-02
Reason
The recall was initiated after the firm discovered through a customer complaint that the Locator Abutments for 5.7 Screw-vent & Compatibles do not fit properly in all sizes of the Zimmer Dental 5.7mm diameter Tapered Screw-Vent Implants (TSV6B8, TSV6B10, TSV6B11, TSV6B13, TSV6B16, TSV6H8, TSV6H10, TSV6H11, TSV6H13 and TSV6H16). The Locator Abutment bottoms out in the pilot hole of these implants c
Root cause
Nonconforming Material/Component
Action
The recall notification letter (Recall Notice) was sent to consignees on April 23,2008. The letter includes the following information on how to handle the recall. It is very important that you determine the use of the Locator Abutment(s) as follows, mark the appropriate box, ,and return this letter to Zest Anchors, Inc. in the stamped, self-addressed envelope: 1. The Locator Abutment(s) listed above WERE NOT used on any Zimmer Dental 5.7mm diameter Tapered Screw-Vent Implants. THE FIT IS CORRECT AND NO ACTION IS NECESSARY. 2. The Locator Abutment(s) listed above WERE used on a Zimmer Dental 5.7mm diameter Tapered Screw-Vent Implant. PLEASE CONTACT OUR DENTAL OFFICE TO ARRANGE THE R PLACEMENT OF THE ABUTMENT WITH A NEWLY DESIGNED LOCATOR ABUTMENT. and 3. We have current STOCK of the Locator Abutment(s) listed above with planned use for a Zimmer Dental 5.7mm diameter Tapered Screw-Vent Implant. PLEASE CONTACT OUR DENTAL OFFICE TO ARRANGE THE REPLACEMENT OF THE ABUTMENT WITH A NEWLY DESIGNED LOCATOR ABUTMENT. For questions or additional information, contact Zest Anchors, Inc. at 800-262-2310 or 760-743-7744.
Quantity
996 units
Distribution
Worldwide Distribution including USA and countries of Spain, Canada, Australia and South Africa.
Lot, serial or model codes
Product Codes: 8190 (1.0 mm), 8191 (2.0 mm), 8192 (3.0mm), 8193 (4.0mm), 8194 (5.0 mm) and 8195 (6.0mm).
510(k) numbers
K994257
PMA numbers
not on file
Product code
DZE
Firm FEI
2023950
Firm city
Escondido
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register