Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Operations, Inc.: Bulk Loader Module for cobas p 512/612 Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis. In vitro diagnostic machine component.

Recall number
Z-0297-2016
Recalling firm
Roche Diagnostics Operations, Inc.
Product
Bulk Loader Module for cobas p 512/612 Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis. In vitro diagnostic machine component.
Status
Terminated
Initiated
2015-09-22
Posted
2015-11-18
Terminated
2016-02-03
Reason
Possible for sample tubes in the loader to be opened and contaminate the system as well as adjacent samples during processing.
Root cause
Device Design
Action
Roche Diagnostics Operations, Inc. sent an Urgent Medical Device Correction letter to affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were asked to retain the Urgent Medical Device Correction letter until a Roche Field Service Representative has installed the BLM modification kit. Consignees were also asked to complete the attached form and fax to 1-877-906-8982, ext 2015. Consignees with questions were instructed to call 1-800-428-2336. For questions regarding this recall call 317-521-3911.
Quantity
1
Distribution
Distributed to TX only.
Lot, serial or model codes
During transport from the Bulk Loader Module (BLM), catalog number 07135645001, to the cobas p 512/p 612 pre-analytical instruments, sample tubes may inadvertently open and spill, causing contamination of the system.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register