FDA Medical Device Recalls (openFDA)
Ion Beam Applications S.A.: IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient Positioning System.
- Recall number
- Z-0295-2024
- Recalling firm
- Ion Beam Applications S.A.
- Product
- IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient Positioning System.
- Status
- Open, Classified
- Initiated
- 2023-09-18
- Posted
- 2023-11-14
- Terminated
- not on file
- Reason
- The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.
- Root cause
- Device Design
- Action
- An Urgent Medical Device Correction issued on October 4, 2023 by email or in-person visit. Waiting for the solution to be deployed, IBA requires the user to only request jog translation motions to recover from a collision situation. IBA also recommends the user to post the notice near the device to remind them of this requirement until the software correction can be deployed. IBA provide an updated Urgent Medical Device Correction notification by November 30, 2023 with a target date for deployment of the software solution to your site. Customers are to complete and return the acknowledgement form to IBA. IBA Contact: Sonia Pinel Head of Post Market Vigilance, Email: Vigilance@iba-group.com, Helpdesk: +32 2 507 20 81 (available 24/7).
- Quantity
- 5 (US), 9 (OUS)
- Distribution
- Worldwide - US Nationwide distribution in the states of Arkansas, Florida, Kansas, and Virginia and the countries of Belgium, France, India, Italy, Netherlands, Spain, Singapore, and Taiwan.
- Lot, serial or model codes
- UDI-DI: (01)05404013801138. Serial numbers (US): SAT133, SAT140, SBF113, SBF128, SBF135 & (OUS): SAT125, SAT127, SAT132, SAT136, SBF107, SBF112, SBF117, SBF124, & SBF125.
- 510(k) numbers
- ["K163500"]
- PMA numbers
- not on file
- Product code
- LHN
- Firm FEI
- 3000256071
- Firm city
- Ottignies-Louvain-La-Neuve
- Firm state
- not on file
- Firm country
- Belgium
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28