Ion Beam Applications S.A.: IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient Positioning System.

Recall number
Z-0295-2024
Recalling firm
Ion Beam Applications S.A.
Product
IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient Positioning System.
Status
Open, Classified
Initiated
2023-09-18
Posted
2023-11-14
Terminated
not on file
Reason
The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.
Root cause
Device Design
Action
An Urgent Medical Device Correction issued on October 4, 2023 by email or in-person visit. Waiting for the solution to be deployed, IBA requires the user to only request jog translation motions to recover from a collision situation. IBA also recommends the user to post the notice near the device to remind them of this requirement until the software correction can be deployed. IBA provide an updated Urgent Medical Device Correction notification by November 30, 2023 with a target date for deployment of the software solution to your site. Customers are to complete and return the acknowledgement form to IBA. IBA Contact: Sonia Pinel Head of Post Market Vigilance, Email: Vigilance@iba-group.com, Helpdesk: +32 2 507 20 81 (available 24/7).
Quantity
5 (US), 9 (OUS)
Distribution
Worldwide - US Nationwide distribution in the states of Arkansas, Florida, Kansas, and Virginia and the countries of Belgium, France, India, Italy, Netherlands, Spain, Singapore, and Taiwan.
Lot, serial or model codes
UDI-DI: (01)05404013801138. Serial numbers (US): SAT133, SAT140, SBF113, SBF128, SBF135 & (OUS): SAT125, SAT127, SAT132, SAT136, SBF107, SBF112, SBF117, SBF124, & SBF125.
510(k) numbers
["K163500"]
PMA numbers
not on file
Product code
LHN
Firm FEI
3000256071
Firm city
Ottignies-Louvain-La-Neuve
Firm state
not on file
Firm country
Belgium

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register