Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. Item Number: VPX4000 (7273)

Recall number
Z-0285-2025
Recalling firm
Zimmer, Inc.
Product
Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. Item Number: VPX4000 (7273)
Status
Open, Classified
Initiated
2024-09-17
Posted
2024-11-01
Terminated
not on file
Reason
A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery
Root cause
Component change control
Action
Zimmer Biomet, in partnership with Saphena Medical, Inc issued Urgent Medical Device Recall Letter- Lot Specific on 9/17/24 via Fed'X. Letter states reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, immediately locate and quarantine affected product in your inventory. 3. Return all affected product from your facility. a. Request a Return Goods Authorization (RGA) number by emailing the completed Attachment 1 Certificate of Acknowledgement form to corporateQuality.PostMarket@zimmerbiomet.com. b. After the RGA number is received, proceed with returning your inventory to the address below: PRIMO MEDICAL GROUP ATTN: ZB Saphena Medical Recall 75 Mill Street Stoughton, MA 02072 c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Mark RECALL on the outside of the returned cartons 4. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 5. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. Upon receipt of affected product, Zimmer Biomet will credit your account. 6. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 7. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
Quantity
432 units
Distribution
Nationwide
Lot, serial or model codes
UDI: +M718VPX40001/$$326051543DD6738Q, +M718VPX4001/$$326062143DD8092J Lot Numbers: 43DD6738, 43DD8092
510(k) numbers
K140603
PMA numbers
not on file
Product code
GEI
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register