FDA Medical Device Recalls (openFDA)
Intuitive Surgical, Inc.: da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 00886874114193) is included in the IS 4000 Vision Cart, Model Number: 381121-31 (UDI 00886874110898) - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.
- Recall number
- Z-0283-2021
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 00886874114193) is included in the IS 4000 Vision Cart, Model Number: 381121-31 (UDI 00886874110898) - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.
- Status
- Terminated
- Initiated
- 2019-06-17
- Posted
- not on file
- Terminated
- 2022-07-18
- Reason
- One da Vinci Xi Endoscope Controller in the field was improperly calibrated during servicing. The result of this improper calibration may result in the tip of the endoscope may reach higher temperature causing the potential for thermal tissue injury and/or overly bright images.
- Root cause
- Equipment maintenance
- Action
- Send formal customer letter communication to the affected site. The firm reported that a representative scheduled an on-site visit to replace the affected unit.
- Quantity
- 1 system
- Distribution
- US Nationwide distribution including in the state of Washington.
- Lot, serial or model codes
- IS4000 Endoscope Controller, Serial Number: 432242 installed on the da Vinci Surgical System, Serial Number: SK2137
- 510(k) numbers
- ["K131861"]
- PMA numbers
- not on file
- Product code
- NAY
- Firm FEI
- 3001675293
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28