FDA Medical Device Recalls (openFDA)
Randox Laboratories Ltd.: Liquid Protein Calibrators (SP CAL (LIQ)) Catalogue Number: IT2691 GTIN: 05055273204032 Product Usage: Liquid Protein Calibrators are intended for in vitro diagnostic use in the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin assays on Clinical Chemistry and Immunoassay systems, Abbott Spectrum, Abbott Aeroset, Abbott Architect i2000, Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems
- Recall number
- Z-0283-2020
- Recalling firm
- Randox Laboratories Ltd.
- Product
- Liquid Protein Calibrators (SP CAL (LIQ)) Catalogue Number: IT2691 GTIN: 05055273204032 Product Usage: Liquid Protein Calibrators are intended for in vitro diagnostic use in the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin assays on Clinical Chemistry and Immunoassay systems, Abbott Spectrum, Abbott Aeroset, Abbott Architect i2000, Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems
- Status
- Terminated
- Initiated
- 2019-08-09
- Posted
- not on file
- Terminated
- 2021-01-08
- Reason
- The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.
- Root cause
- Nonconforming Material/Component
- Action
- 1. Review your reagent inventory and replace the value sheet with the new values provided. 2. Review results generated with the affected batches in line with the clinical profile of the patient. 3. Discuss the contents of this notice with your Medical Director. 4. Complete and return the response form 12187-QA to teclhnical.services@randox.com within five working days.
- Quantity
- 100 kits
- Distribution
- US Nationwide distribution in the states of WV, ME, NJ, CA, and Puerto Rico.
- Lot, serial or model codes
- Batch/Lot # 440973 /1748 1752IT 451414 /1748 1752IT 470531 /1813 1817IT 480859 /1813 1817IT
- 510(k) numbers
- ["K061056"]
- PMA numbers
- not on file
- Product code
- JIX
- Firm FEI
- 1000361607
- Firm city
- Crumlin
- Firm state
- North
- Firm country
- Ireland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28