Skip to the content

FDA Medical Device Recalls (openFDA)

Chembio Diagnostic System Inc: OTC Pregnancy Tests. Name: Sure Check hCG Cassette Pregnancy Test. --- Description: the Sure Check hCG Cassette Pregnancy test consists of a lateral flow membrane pad containing anti- hCG monoclonal antibody conjugated to colloidal gold contained in a plastic cassette. It differs from the Sure Check Midstream product in that the lateral flow membrane pad is placed in a cassette and the urine sample is added by pipette rather than by urinating on the absorbent wick. The device is packaged in a foil pouch with desiccant. This product is shipped as bulk packaged product of varying quantities (Catalog # PT127-OTC ordered by Innovative Chemical Co., and Pruebas Moderna Del Istmo; Catalog # PT210 ordered by LABSCO). --- 510(k) K933529/A.

Recall number
Z-0277-05
Recalling firm
Chembio Diagnostic System Inc
Product
OTC Pregnancy Tests. Name: Sure Check hCG Cassette Pregnancy Test. --- Description: the Sure Check hCG Cassette Pregnancy test consists of a lateral flow membrane pad containing anti- hCG monoclonal antibody conjugated to colloidal gold contained in a plastic cassette. It differs from the Sure Check Midstream product in that the lateral flow membrane pad is placed in a cassette and the urine sample is added by pipette rather than by urinating on the absorbent wick. The device is packaged in a foil pouch with desiccant. This product is shipped as bulk packaged product of varying quantities (Catalog # PT127-OTC ordered by Innovative Chemical Co., and Pruebas Moderna Del Istmo; Catalog # PT210 ordered by LABSCO). --- 510(k) K933529/A.
Status
Terminated
Initiated
2004-10-06
Posted
2004-12-03
Terminated
2005-03-24
Reason
Poor sealing of pouches and degradation of products. Firm received an e-mail from a distributor of an increase in complaints. In response to the complaints, the distributor conducted its own functional testing and noticed a degradation in performance of the two lots. Chembio's investigation revealed invalid results (no control lines) and some false negatives.
Root cause
Other
Action
Recall letters on 10/6/04, followed by a second letter on/about 10/20/04. Recall letters mailed on 11/9/04 to cover recall expansion for 4 additional lots.
Quantity
42,005 (total of 2 lots)
Distribution
Distributors in KS, TN, NC, MA, NY, KY, FL, NV, and Puerto Rico for further distribution to their retail customers. Products also distributed to accounts in Barbados WI, Israel, Costa Rica and Canada.
Lot, serial or model codes
Lot U040904 (Exp. 30 SEP 06) ; Recall Expansion: lot U032904/1 (Exp. 30 AUG 06 - 14,364 units distributed)
510(k) numbers
["K933529"]
PMA numbers
not on file
Product code
LCX
Firm FEI
2431980
Firm city
Medford
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register