Skip to the content

FDA Medical Device Recalls (openFDA)

Encore Medical, LP: DJO EMPOWR Knee Punch Handle, REF: 801-05-040

Recall number
Z-0258-2024
Recalling firm
Encore Medical, LP
Product
DJO EMPOWR Knee Punch Handle, REF: 801-05-040
Status
Open, Classified
Initiated
2023-09-29
Posted
2023-11-07
Terminated
not on file
Reason
Knee punch handle tip is breaking during surgery due to the lack of heat treatment, which may lead to a disruption in surgery.
Root cause
Process change control
Action
On 9/29/23, recall notices were emailed to customers who were asked to do the following: 1) Knee punch handles should be visually inspected prior to use for any damage. Upon completion of surgery, visually inspect and verify there is no breakage. The attached quality bulletin provides guidance on an acceptable and unacceptable device. 2) Contact your customer service rep at customerservice@enovis.com to report failures or to request replacements. 3) Complete and return the response form that asks customers to check all related devices in inventory and report the number of related devices in inventory. Questions can be directed to livia.kuruvila@enovis.com
Quantity
406
Distribution
US: VA, TN, MN, TX, IL, IN, KY, OH, SC, MS, RI, ME, NJ, NY, MD, CA, KS, LA, PR, AL, FL, GA, AZ, WA
Lot, serial or model codes
UDI-DI: 00190446238881, Lots: 305197L02D, 305197L02E, 318871L10D, 318871L10E, 305197L02B, 305197L02C, 318871L10B, 318871L10C, 325045L17A, 325045L17B, 325045L17C, 305197L02, 318871L10, 325045L17, 327669L03, 331146L01, 332158L06, 334314L04, 336735L04, 336801L01, 340391L01, 341961L01, 342586L01, 346101L01, 350076L01, 351106L02, 360277L01, 366833L03, 374648L01, 380510L01
510(k) numbers
["K171991"]
PMA numbers
not on file
Product code
MBH
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register