Fujifilm Medical Systems U.S.A., Inc.: FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations.

Recall number
Z-0232-2018
Recalling firm
Fujifilm Medical Systems U.S.A., Inc.
Product
FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations.
Status
Terminated
Initiated
2017-06-15
Posted
not on file
Terminated
2018-05-09
Reason
A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.
Root cause
Device Design
Action
FUJIFILM Medical Systems U.S. A. Inc. (FMSU) sent "Urgent: Medial Device Correction" recall notification letter dated June 21, 2017 to all Customers. Customers were notified of the recall, potential issue, potential hazard, description of the issue, the actions planned to correct the issue, the actions to be taken by the customer and contact information. A Customer Feedback Form was provided accompanying the letter. FMSU service personnel will contact all of the medical facilities where the applicable products have been installed to arrange for this correction and visit to take corrective measures. As FMSUs planned field correction at the affected sites, FMSU will install the additional protective parts preventing the X-ray tube assembly from coming into contact with the patient or operator should separation of the tube mounting flange from the tube holding shaft occur. For further questions, please call (203) 324-2000.
Quantity
3
Distribution
US Distribution.
Lot, serial or model codes
UDI: (01)04540217052226(21)MP95A9F6B001; (01)04540217052226(21)MP95A9F6A001; (01)04540217052226(21)MP95A8F5A001.  Serial Numbers: FDR Visionary Suite System S/N MP95A9F6B001 MP95A9F6A001 MP95A8F5A001
510(k) numbers
K152294
PMA numbers
not on file
Product code
KPR
Firm FEI
1000513161
Firm city
Stamford
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register