Physio-Control, Inc.: LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o 99577-001256 o 99577-001368 o 99577-001955 o 99577-001956 o 99577-001957 o 99577-001958 o 99577-001962 LIFEPAK 15 intended use monitor/defibrillator.
- Recall number
- Z-0230-2025
- Recalling firm
- Physio-Control, Inc.
- Product
- LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o 99577-001256 o 99577-001368 o 99577-001955 o 99577-001956 o 99577-001957 o 99577-001958 o 99577-001962 LIFEPAK 15 intended use monitor/defibrillator.
- Status
- Open, Classified
- Initiated
- 2024-08-30
- Posted
- 2024-10-25
- Terminated
- not on file
- Reason
- Due to an out of tolerance tool being used on monitor/defibrillator systems.
- Root cause
- Vendor change control
- Action
- On 09/30/2024, the firm mailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them that Stryker discovered that an out of tolerance service tool was used on affected LP15 devices in which there is a potential for Noninvasive Blood Pressure (NIBP), System or Auxiliary Power Connectors to be loose. There is a possibility that water or dust may accumulate in the ingress and can lead to permanent or intermittent failures. Customers are instructed to: 1. Immediately check internal inventory to locate the product listed on the attached business reply form. 2. To confirm receipt of this Medical Device Correction Notice and understanding of the provided information, email the enclosed Business Reply Form (BRF) by October 31st, 2024 date, to RSRecall@Stryker.com. 3. Upon receipt of the completed business reply form, Stryker will contact customers to arrange for corrective action on the device. 4. There are no additional actions that users can and should take. Please continue to follow the Operating Instructions and perform the daily checks and user tests. These tests can be found in the chapter Maintaining the Equipment subsection General Maintenance and Testing . 5. In the interim, Stryker recommends that you continue to use your LP15 device. 6. Maintain awareness of this communication internally until all required actions have been completed within your facility. Please ensure this letter is kept with the affected device until the correction has been completed. 7. Inform Stryker if any of the subject devices have been distributed to other organizations by contacting Stryker at: a. Device Tracking: rsdevicetracking@stryker.com b. Website: https://www.strykeremergencycare.com/service-support-overview/device-registration/ If you have any questions or concerns, please contact Customer Service +1 800 442 1142 or by email RSRecall@Stryker.com.
- Quantity
- 105 systems
- Distribution
- U.S.: CT, GA, MA, NY, OH, WA, and WV. O.U.S.: N/A
- Lot, serial or model codes
- LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o 99577-001256 o 99577-001368 o 99577-001955 o 99577-001956 o 99577-001957 o 99577-001958 o 99577-001962 UDI-DI codes: o 00883873808214 o 00883873988022 o 00883873979426 o 00883873979358 o 00883873911693 o 00883873979198 o 00883873911648 o 00883873884386 o 00883873871713 o 00883873871706 o 00883873871690 o 00883873871683 o 00883873911631 o 00883873871645 Serial Numbers: o 45435833 o 46470647 o 46686048 o 46686573 o 46686708 o 46687360 o 46906073 o 49037777 o 49704552 o 38166647 o 38282988 o 38290136 o 38291331 o 38802782 o 39541663 o 39541666 o 39752193 o 40259639 o 41446867 o 41447020 o 41453331 o 42325059 o 43574409 o 43984215 o 45040678 o 45308104 o 45455023 o 46017745 o 46572054 o 46576300 o 48349664 o 48605645 o 48605663 o 49001744 o 49025185 o 49070762 o 49087040 o 49087053 o 49224619 o 49374174 o 49422314 o 49619921 o 38200440 o 38805564 o 38805583 o 39382735 o 41446693 o 41446882 o 41446975 o 41446977 o 41447027 o 41453239 o 41893891 o 42324176 o 42610924 o 43181935 o 43182268 o 43182389 o 43462227 o 43465372 o 43487806 o 43610792 o 43791346 o 43841374 o 43860850 o 43862247 o 43869747 o 44169862 o 44356379 o 44360003 o 44601246 o 45590467 o 45907317 o 46364496 o 46691551 o 47447195 o 47515791 o 47539203 o 47539310 o 47785570 o 47794250 o 48524850 o 48583111 o 48585709 o 48600317 o 48605407 o 48605467 o 48704914 o 48749443 o 49022830 o 49031342 o 49036187 o 49056308 o 49057928 o 49123367 o 49126651 o 49127012 o 49132399 o 49132740 o 49134638 o 49511402 o 49513814
- 510(k) numbers
- not on file
- PMA numbers
- ["P160026"]
- Product code
- MKJ
- Firm FEI
- 3015876
- Firm city
- Redmond
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28