Medacta Usa: The MectaLIF Oblique Handle, model number 03.22.10.0262, is used to insert the MectaLIF oblique spinal cage implants.

Recall number
Z-0225-2015
Recalling firm
Medacta Usa
Product
The MectaLIF Oblique Handle, model number 03.22.10.0262, is used to insert the MectaLIF oblique spinal cage implants.
Status
Terminated
Initiated
2014-10-21
Posted
2014-11-18
Terminated
2016-03-14
Reason
The MectaLIF Oblique Handle has the laser marking "MEDIAL" on the incorrect side and could result in incorrect insertion of the device during surgery.
Root cause
Employee error
Action
Medacta sent an URGENT MEDICAL DEVICE RECALL letter dated October 21, 2014 were via e-mail to all sales representatives and distributors. The letters included instructions for the sales represetantives and distributors to: 1) check their inventory for the recalled devices; 2) immedaitely sequester the recalled devices and do not use them during surgery; 3) if the devices have been further distributed, inform them of the recall; and, 4) replacement devices will be provided. Customers with questions can contact Medacta USA at 805-910-6511.
Quantity
32 oblique handles
Distribution
Worldwide Distribution - US including CA, CO, ID and Internationally to Australia and Germany.
Lot, serial or model codes
Model Number: 03.22.10.0262;  Lot Number: 1314256
510(k) numbers
["K110927"]
PMA numbers
not on file
Product code
MAX
Firm FEI
3015500078
Firm city
Chicago
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register