Skip to the content

FDA Medical Device Recalls (openFDA)

Spectrum Medical Ltd.: Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and arterial circuit tubing. Catalog Numbers: a) 51-000013-00, b) 51-000008-00

Recall number
Z-0224-2025
Recalling firm
Spectrum Medical Ltd.
Product
Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and arterial circuit tubing. Catalog Numbers: a) 51-000013-00, b) 51-000008-00
Status
Completed
Initiated
2023-03-13
Posted
2024-10-24
Terminated
not on file
Reason
The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock.
Root cause
Component design/selection
Action
All potentially affected product was identified. The firm corrected all products between 03/13/2024 and 08/07/2024 during service visits.
Quantity
172 units
Distribution
US Nationwide distribution including in the states of Alabama, California, Florida, Kentucky, Maine, Massachusetts, Minnesota, Missouri, New Jersey, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, West Virginia.
Lot, serial or model codes
UDI/DI 05060434420770: Serial Numbers: a) AU6004040, AU6004041, AU6004042, AU6004043, AU6004044, AU6004046, AU6004050, AU6004051, AU6004052, AU6004053, AU6004054, AU6004056, AU6004057, AU6004058, AU6004060, AU6004061, AU6004062, AU6004063, AU6004064, AU6004065, AU6004066, AU6004069, AU6004070, AU6004071, AU6004072, AU6004073, AU6004074, AU6004075, AU6004136, AU6004137, AU6004138, AU6004139, AU6004140, AU6004141, AU6004142, AU6004143, AU6004144, AU6004145, AU6004146, AU6004147, AU6004148, AU6004159, AU6004160, AU6004161, AU6004162, AU6004167, AU6004168, AU6004169, AU6004170, AU6004171, AU6004172, AU6004173, AU6004174, AU6004175, AU6004176, AU6004177, AU6004178, AU6004179, AU6004180, AU6004181, AU6004182, AU6004183, AU6004184, AU6004185, AU6004186, AU6004187, AU6004188, AU6004195, AU6004197, AU6004198, AU6004200, AU6004201, AU6004202, AU6004203, AU6004204, AU6004205, AU6004207, AU6004332, AU6004333, AU6004348, AU6004349, AU6004350, AU6004352, AU6004353, AU6004354, AU6004532, AU6004577, AU6004578, AU6004579, AU6004580, AU6004587, AU6004588, AU6004590, AU6004591, AU6004603, AU6004604, AU6004605, AU6004606, AU6004607, AU6004609, AU6004610, AU6004613, AU6004731, AU6004771, AU6004773, AU6004776, AU6004777, AU6004778, AU6004779, AU6004781, AU6004782, AU6004919, AU6004920, AU6004921, AU6004922, AU6004923, AU6004924, AU6004925, AU6004926, AU6004927, AU6004929, AU6004930, AU6004955, AU6004957, AU6004958, AU6004960, AU6004963, AU6004966, AU6004968, AU6004970, AU6004971, AU6004972, AU6005010, AU6005021, AU6005022, AU6005023, AU6005024, AU6005025, AU6005026, AU6005027, AU6005028, AU6005029, AU6005030, AU6005037, AU6005038, AU6005039, AU6005040, AU6005041, AU6005042, AU6005057, AU6005065, AU6005109, AU6005110, AU6005118, AU6005124, AU6005125, AU6005355, AU6006351; Serial Numbers: b) AU8001273, AU8001275, AU8001346, AU8001347, AU8001348, AU8001359, AU8001360, AU8001366, AU8001403, AU8001406, AU8001407, AU8001446, AU8001447, AU8001449
510(k) numbers
["K173834"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
3006073153
Firm city
Gloucester
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register