Nihon Kohden America Inc: Nihon Kohden Electric Stimulator, Model number: MS-210BK (Optional accessories of MEB-2200A -Neuropack Evoked Potential and EMG Measuring System)

Recall number
Z-0218-2007
Recalling firm
Nihon Kohden America Inc
Product
Nihon Kohden Electric Stimulator, Model number: MS-210BK (Optional accessories of MEB-2200A -Neuropack Evoked Potential and EMG Measuring System)
Status
Terminated
Initiated
2006-08-23
Posted
2006-11-28
Terminated
2008-05-23
Reason
When the device is subject to physical shock, such as a drop to the floor, it may output a different value from the preset value. This may result in a superficial burn on the patient if the stimulation is done for a long time.
Root cause
Other
Action
All customers will be called by phone, then faxed the Product Notification starting 09/01/2006 Customers were instructed to indicate the serial numbers on each of your MS-210BK stimulators and fax the MS-210BK PRODUCT RECALLREPLACEMENT letter back to Nihon Kohden at: 949-580-15550 or email to customerservice@nkusa.com. Upon receipt of this letter, Nihon Kohden''s customer service department will send a replacement unit.
Quantity
97 units in the U.S.
Distribution
Nationwide.
Lot, serial or model codes
Any serial numbers up to 02160
510(k) numbers
["K991899"]
PMA numbers
not on file
Product code
GWF
Firm FEI
2080783
Firm city
Foothill Ranch
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register