Skip to the content

FDA Medical Device Recalls (openFDA)

ELITech Clinical Systems SAS: Envoy 500 system includes the following product reference numbers and names: 55345 ENVOY 500 Glucose 55450 ENVOY 500 URIC ACID REAGENT KIT 55435 ENVOY 500 TRIGLYCERIDES REAGENT KIT, 55297 ENVOY 500 CHOLESTEROL REAGENT KIT, 55335 ENVOY 500 CREATININE REAGENT KIT 55301 ENVOY 500 HDL CHOLESTEROL REAGENT 55310 ENVOY 500 LDL CHOLESTEROL REAGENT KIT

Recall number
Z-0217-2016
Recalling firm
ELITech Clinical Systems SAS
Product
Envoy 500 system includes the following product reference numbers and names: 55345 ENVOY 500 Glucose 55450 ENVOY 500 URIC ACID REAGENT KIT 55435 ENVOY 500 TRIGLYCERIDES REAGENT KIT, 55297 ENVOY 500 CHOLESTEROL REAGENT KIT, 55335 ENVOY 500 CREATININE REAGENT KIT 55301 ENVOY 500 HDL CHOLESTEROL REAGENT 55310 ENVOY 500 LDL CHOLESTEROL REAGENT KIT
Status
Terminated
Initiated
2015-09-22
Posted
2015-11-05
Terminated
2016-03-04
Reason
ELITech clinical Systems reagents based on Trinder reaction used on the Envoy 500 instrument system are recalled due to potential drug interference. No complaints were received.
Root cause
Under Investigation by firm
Action
The firm sent recall letters or Technical Bulletins to US consignees on 9/22/15 via registered mail. Technical Bulletin #1022 (for EON 100 users) Technical Bulletin #2008 (for EON 300 users) Technical Bulletin #5064 (for ENVOY users) Technical Bulletin #3009 (for Selectra users) ELITechGroup is not requesting return of the affected product. The method sheets and package inserts will be updated with the new interference claims. When available, the revised labeling will be provided to the customer for customer's use until the new IFU is included in the product packaging. Consignees are requested to return the Confirmation of Receipt form back to the firm by FAX to 401-642-9001 or send email to customercare@elitechgroup.com or send via US Post: Vital Diagnostics, 27 Wellington Rd., Lincoln, Rl. If you have any questions, please contact Vital Diagnostics Technical Support at 1-800-345-2822 or 855-ELITECH
Quantity
13455 systems
Distribution
US Nationwide Distribution
Lot, serial or model codes
all lots
510(k) numbers
K100525
PMA numbers
not on file
Product code
CGA
Firm FEI
3007662974
Firm city
SEES
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register