Wipro GE Healthcare Private Ltd.: GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube

Recall number
Z-0213-2026
Recalling firm
Wipro GE Healthcare Private Ltd.
Product
GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube
Status
Open, Classified
Initiated
2025-09-05
Posted
2025-10-21
Terminated
not on file
Reason
Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
Root cause
Software design
Action
GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 09/05/2025, by letter delivered via traceable means. The notice explained the issues and requested the following: "Actions to be taken by Customer/User Pending correction by GE HealthCare, you can continue to use your device by following the instructions below. 1. Determine the number of central stations that are connected to the network including those in any offsite Central Monitoring Unit. 2. Ensure the number of connected central stations are fewer than 118 or more than 120. If you need assistance completing steps 1 or 2, please contact GE HealthCare Service or your local Service Representative. Note: Any versions prior to CSCS V3.0.5 are not affected by this issue. Ensure all users are aware of, understand, and follow these instructions. Please retain this document for your records. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity
173 units
Distribution
Worldwide distribution.
Lot, serial or model codes
GTIN 00195278343512, Serial Numbers: STU25210418SA, STU25210417SA, STU25210419SA, STU25250440SA, STU25250439SA, STU25260447SA, STU25210407SA, STU25210406SA, STU24200003SA, STU24200006SA, STU24200005SA, STU24200004SA, STU23240015SA, STU24240019SA, STU24240020SA, STU24240007SA, STU24240022SA, STU24240021SA, STU24330020SA, STU24240014SA, STU24240013SA, STU24330019SA, STU24330022SA, STU24330016SA, STU24240005SA, STU24240008SA, STU24240012SA, STU24330018SA, STU24330021SA, STU24330017SA, STU23160011SA, STU23240007SA, STU24310001SA, STU25210433SA, STU25210432SA, STU25200390SA, STU25200392SA, STU25200391SA, STU25200399SA, STU25200397SA, STU25200395SA, STU25200393SA, STU25200398SA, STU25200396SA, STU25250438SA, STU25270475SA, STU25270456SA, STU25270468SA, STU25270458SA, STU25270466SA, STU25270476SA, STU25270469SA, STU25270473SA, STU25270459SA, STU25270474SA, STU25270467SA, STU25210430SA, STU25210429SA, STU25180343SA, STU25190347SA, STU25190348SA, STU24200007SA, STU25190360SA, STU25240436SA, STU25190361SA, STU25190362SA, STU25190363SA, STU25240437SA, STU25190359SA, STU24510027SA, STU24510032SA, STU24510010SA, STU24510020SA, STU24510017SA, STU24510026SA, STU24510018SA, STU24510024SA, STU24510025SA, STU24510031SA, STU24510011SA, STU24510029SA, STU24510023SA, STU24510030SA, STU24510021SA, STU24510019SA, STU24510022SA, STU25190350SA, STU25200388SA, STU25200389SA, STU25210412SA, STU25190349SA, STU25190381SA, STU25190376SA, STU25190383SA, STU25190374SA, STU25190367SA, STU25190372SA, STU25190379SA, STU25190368SA, STU25190352SA, STU25190358SA, STU25190370SA, STU25190366SA, STU25190355SA, STU25190354SA, STU25190382SA, STU25190380SA, STU25190371SA, STU25190377SA, STU25190353SA, STU25190351SA, STU25190364SA, STU25190357SA, STU25190384SA, STU25190369SA, STU25190373SA, STU25190356SA, STU25190365SA, STU25190378SA, STU25190385SA, STU25210410SA, STU25210411SA, STU25210409SA, STU25230435SA, STU25180346SA, STU25180344SA, STU25180345SA, STU25210413SA, STU25210414SA, STU25210423SA, STU25210431SA, STU25210415SA, STU25270453SA, STU25260450SA, STU25260451SA, STU25290488SA, STU25210408SA, STU25210400SA, STU25270454SA, STU25210404SA, STU25210421SA, STU25270464SA, STU25210420SA, STU25270462SA, STU25210425SA, STU25270461SA, STU25210426SA, STU25210428SA, STU25210427SA, STU25270463SA, STU25270460SA, STU24460058SA, STU24460059SA, STU24460071SA, STU24460060SA, STU25200387SA, STU25210401SA, STU25200386SA, STU25210424SA, STU25270457SA, STU25270465SA, STU25300492SA, STU25210402SA, STU25210403SA, STU25250446SA, STU25210422SA, STU25210405SA, STU25260449SA, STU25290481SA, STU25290482SA, STU25290483SA, STU25290484SA, STU25210416SA.
510(k) numbers
K162012
PMA numbers
not on file
Product code
DXJ
Firm FEI
3007734888
Firm city
Bengaluru
Firm state
not on file
Firm country
India

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register