Skip to the content

FDA Medical Device Recalls (openFDA)

Biosense Webster, Inc.: SMARTABLATE RF System Generator

Recall number
Z-0210-2018
Recalling firm
Biosense Webster, Inc.
Product
SMARTABLATE RF System Generator
Status
Terminated
Initiated
2017-06-14
Posted
not on file
Terminated
2021-01-11
Reason
Generator may inadvertently deliver RF energy if the foot pedal is placed in close proximity and oriented toward the Stereotaxis magnets.
Root cause
Labeling design
Action
All new U.S. and OUS customers, who have purchased the SMARTABLATE" Generator system since the launch of the original field action in March 2015 to the time the IFU was updated in November 2015, will be notified. The previously distributed Field Safety Notification Letter for the SMARTABLATE" Generator (Part Numbers# M490006 and M490007) and customer Acknowledgement Form will be sent to these new customers. Instructions included review the Field Safety Notice carefully, provide the notice to appropriate personnel in facility, ensure the SMARTABLATE System Foot Pedal is not located within the Stereotaxis RMT Lab, complete and return the Acknowledgement Forms, and maintain awareness of the notice.
Quantity
185 units total
Distribution
Distributed to the following states: CA, CO, GA, IL, IN, MI, NM, NY, OH, PA, TN, TX, UT, WA, WI. Product was also distributed to the following countries: Germany, France, Austria, Sweden Italy, Belgium, Denmark, and Singapore.
Lot, serial or model codes
Product Codes M490006, M490007
510(k) numbers
not on file
PMA numbers
["P990071S017"]
Product code
LPB
Firm FEI
3003113629
Firm city
Irwindale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register