Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Artis zeego, Fluoroscopic X-Ray System Models: 10280959,10502505, 10848283 in combination with the following workplace/system application software models: Product/Model #: syngo X Workplace, 10281061 syngo X Workplace, 10281299 syngo X Workplace, 10502522 Syngo Application Software, 10848815

Recall number
Z-0200-2022
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Artis zeego, Fluoroscopic X-Ray System Models: 10280959,10502505, 10848283 in combination with the following workplace/system application software models: Product/Model #: syngo X Workplace, 10281061 syngo X Workplace, 10281299 syngo X Workplace, 10502522 Syngo Application Software, 10848815
Status
Open, Classified
Initiated
2021-09-24
Posted
not on file
Terminated
not on file
Reason
Software error, the software can produce an incorrect interpretation of the table rotation if the table has been rotated by more than 4degrees in either direction. This can cause the 3D image reconstruction to be shown rotated by up to 5 degrees relative to the patient and to be used inappropriately by the system and result in incorrect treatment of the patient
Root cause
Software design
Action
Siemens Healthineers issued Urgent Medical Device Correction Letter (AX059/21/S) dated 9/24/21. Letter states reason for recall, health risk and action to take: To avoid possible incorrect patient treatment, Siemens recommends a working position with 0 degree table rotation be used if intraoperative 3D images are required for subsequent steps (e.g., image superimposition, C-arm positioning, laser cross). If it should still be necessary to use a table rotation other than 0 degrees for intraoperative 3D images, please verify the C-arm position, needle position and image superimposition, by means of fluoroscopy. Siemens will provide a software update for all affected systems in which the system calculates table rotations and 3D reconstructions correctly. This corrective action will be distributed via Update Instruction AX058/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein.
Quantity
43 US
Distribution
US Nationwide distribution.
Lot, serial or model codes
Artis zeego systems with Artis patient table and AppSW VD20C/VD20N
510(k) numbers
K181407
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register