American Contract Systems Inc.: Medical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 PERP25AQ PEI RETINAL PACK-206025 RFEY50AC EYE PACK-206026 UIVT47AS VITRECTOMY SUPPLY

Recall number
Z-0185-2026
Recalling firm
American Contract Systems Inc.
Product
Medical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 PERP25AQ PEI RETINAL PACK-206025 RFEY50AC EYE PACK-206026 UIVT47AS VITRECTOMY SUPPLY
Status
Open, Classified
Initiated
2025-09-04
Posted
2025-10-16
Terminated
not on file
Reason
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Root cause
Nonconforming Material/Component
Action
Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.
Quantity
32,433 total
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Lot, serial or model codes
BFCT04X UDI-DI 191072225955 Lot 7992111 bag serial number N/A; CMEY17AI UDI-DI 191072211514 lot 8082511 bag serial numbers 17390960 17390970 17391031; PERP25AQ UDI-DI 191072199720 lot 8078511 bag serial number 17507000; RFEY50AC UDI-DI 191072202918 lot 8048811 bag serial number 17483686; UIVT47AS UDI-DI 191072224569 lot 7969711 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected
510(k) numbers
not on file
PMA numbers
not on file
Product code
OJK
Firm FEI
3007069947
Firm city
Grimes
Firm state
IA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register