Siemens Healthcare Diagnostics, Inc: Siemens ADVIA Centaur BR Assay: Catalog Number: US: REF 02419937 (10214722) (50 Tests); REF 03896216(10334837)(250 Tests) Not CE Marked; OUS: REF03204829(10310271)(50 Tests); REF 01157807(1 0314722) (250 Tests) CE Marked; REF 09010686 (10340081)(250 Tests). The ADVIA Centaur BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur and ADVIA Centaur XP systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage Ill breast cancer. Serial testing for CA 27.29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression or regression of disease in response to treatment.

Recall number
Z-0162-2013
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
Siemens ADVIA Centaur BR Assay: Catalog Number: US: REF 02419937 (10214722) (50 Tests); REF 03896216(10334837)(250 Tests) Not CE Marked; OUS: REF03204829(10310271)(50 Tests); REF 01157807(1 0314722) (250 Tests) CE Marked; REF 09010686 (10340081)(250 Tests). The ADVIA Centaur BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur and ADVIA Centaur XP systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage Ill breast cancer. Serial testing for CA 27.29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression or regression of disease in response to treatment.
Status
Terminated
Initiated
2012-09-11
Posted
2012-10-28
Terminated
2016-09-01
Reason
BR Assay for CA 27.29 Calibration Interval Change due quality control (QC) results outside of the expected ranges.
Root cause
Under Investigation by firm
Action
Siemens sent two "URGENT FIELD SAFETY NOTICE" letters dated August and September 2012 to all affected customers. The letters identified the product, problem, and actions to be taken by the customers. A Field Correction Effectiveness Check Form was included for customers to complete and return.
Quantity
15,455 kits
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Australia, Brazil, Canada, Columbia, Japan, Korea, Mexico, New Zealand, Peru, Saudi Arabia, and Spain.
Lot, serial or model codes
BR Assay for CA 27.29 BR Assay Kit Lots ReadyPack¿ Lot Number Ending in 198:. 89291198,86469198,85752198,88939198, 87757198, 85459198, 88319198, 86295198, 85194198  ReadyPack¿ Lot Number Ending in 202:  93573202, 91045202, 93664202, 90880202, 93157202,91126202,90494202.  BR Assay for CA 15-3 ReadyPack BR Assay Kit Lots Lot Number Ending in 199:  87425199,85862199,86949199,85945199
510(k) numbers
["K982680"]
PMA numbers
not on file
Product code
MOI
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register