Skip to the content

FDA Medical Device Recalls (openFDA)

First Medical Source LLC: Medpro Elastomeric Infusion Pump. AccuFlux, Model # CT-0020-100H. Product Usage: The devices are intended for patients requiring intravenous, percutaneous, subcutaneous, intra-operative sites or epidural administration of medications. It is the responsibility of the user to ensure that the medication is prepared and administered accordance with the drug manufacturers package insert. The devices deliver controlled amounts of medication directly to the intraoperative site for pain management and or antibiotic administration. The devices infuse the medication at an hourly flow rate. Medications are infused intraoperatively and postoperatively through intramuscular or subcutaneous routes. The devices are also intended for controlled delivery of local anesthetics in close proximity to nerves for post operative regional anesthesia and pain management. Routes of administration may be intraoperative or percutaneous. It is for continuous infusion of medications for general infusion use, including antibiotic delivery and chemotherapy.

Recall number
Z-0161-2013
Recalling firm
First Medical Source LLC
Product
Medpro Elastomeric Infusion Pump. AccuFlux, Model # CT-0020-100H. Product Usage: The devices are intended for patients requiring intravenous, percutaneous, subcutaneous, intra-operative sites or epidural administration of medications. It is the responsibility of the user to ensure that the medication is prepared and administered accordance with the drug manufacturers package insert. The devices deliver controlled amounts of medication directly to the intraoperative site for pain management and or antibiotic administration. The devices infuse the medication at an hourly flow rate. Medications are infused intraoperatively and postoperatively through intramuscular or subcutaneous routes. The devices are also intended for controlled delivery of local anesthetics in close proximity to nerves for post operative regional anesthesia and pain management. Routes of administration may be intraoperative or percutaneous. It is for continuous infusion of medications for general infusion use, including antibiotic delivery and chemotherapy.
Status
Terminated
Initiated
2011-07-09
Posted
2012-10-30
Terminated
2014-09-22
Reason
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being reclassified by the Agency as a Class I. The recall was initiated because First Medical Source has confirmed that these lots may have a higher flow rate than specified. The use of this product may lead to over-administration of drug solutions to the patients. The pr
Root cause
Nonconforming Material/Component
Action
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being reclassified by the Agency as a Class I. First Medical Source received an email from Medpro International with the following lnstruction to Customer - Informed customers to remove product from sales/discontinue use, cease distribution, remove the products from inventory and quarantine, inform their customers to retum the unused products to them, and destroy the product. Distributors' customer service will contact the customer to coordinate the return of the goods in question where required. First Medical Source emailed to their customer, Williams Medical Technologies, to informed them of the recall of the Medpro Elastomeric infusion pumps. Customer was instructed to inform them of the date and method of disposal of recalled products.
Quantity
500 units
Distribution
US Nationwide distribution in the state of IL
Lot, serial or model codes
AccuFlux, Lot#: 91209.
510(k) numbers
["K081905"]
PMA numbers
not on file
Product code
MEB
Firm FEI
3007008418
Firm city
Laguna Beach
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register