Sorin Group Deutschland GmbH: S5 Heart-lung machine, Stockert S5 System. Catalogue No: 48-30/40/50-00. The Stockert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six(6) hours or less.

Recall number
Z-0142-2014
Recalling firm
Sorin Group Deutschland GmbH
Product
S5 Heart-lung machine, Stockert S5 System. Catalogue No: 48-30/40/50-00. The Stockert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six(6) hours or less.
Status
Terminated
Initiated
2013-09-24
Posted
2013-11-04
Terminated
2014-07-10
Reason
Sorin Group is recalling certain models of S5 Perfusion Systems due to malfunction of cardioplegia control.
Root cause
Software Manufacturing/Software Deployment
Action
Sorin Group sent an Urgent Field Safety Notice dated September 24, 2013, to all affected customers. The letter informed users of the potential for failing of automatic cardioplegia delivery function and notified them that they would be contacted by the Sorin Service Group to arrange an appointment to update the firmware on site. Customers were asked to complete the Response form to confirm they had received, read and understood the Field Notice. For questions regarding this recall call 303-467-6527.
Quantity
84
Distribution
Worldwide Distribution - USA including WA, NY, VA, TN, SC, TX, CA, CO, MA, WI, MD, AL, ND, and Puerto Rico and Internationally to Italy, Finland, Germany, Sweden, and Great Britain.
Lot, serial or model codes
Serial Numbers: 48E00235,48E00339,48E00342,48E01753,48E02767,48E02753,48E02745,48E02746,48E02748,48E02416,48E01033,480E1038,48E01040,48E01041,48E01043,48E01044,48E01051,48E02631,48E01183,48E02758,48E02764,48E01832,48E01834,48E02278,48E02722,48E02732,48E02326,48E02797,48E02793,48E02791,48E02781,48E02785,48E02786,48E02783,48E02772,48E02802,48E02778,48E02784,48E02788,48E02789,48E02779,48E02792,48E02790,48E02771,48E02773,48E02776,48E02799,48E02801
510(k) numbers
["K071318"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
3002808267
Firm city
Munchen
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register