GE Medical Systems, LLC: Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System.

Recall number
Z-0138-2010
Recalling firm
GE Medical Systems, LLC
Product
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System.
Status
Terminated
Initiated
2009-04-16
Posted
2009-11-09
Terminated
2015-12-03
Reason
Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room monitor stating that three minutes are left before the X-ray is inhibited. This could lead to a delay in treatment and possible additional X-ray exposure and contra
Root cause
Device Design
Action
Consignees were sent on May 1, 2009 a GE Healthcare " Urgent Medical Device Correction" letter dated April 16, 2009. The letter was addressed to Hospital Administrator/ Risk Manger, Mangers of Radiology/ Cardiology, and Radiologists/ Cardiologists. The letter described the problem, Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Quantity
15 AND 46 UNLOCATED
Distribution
Worldwide distribution: USA, ARGENTINA, BELGIUM, BRAZIL, BULGARIA, CHINA, CROATIA, EGYPT, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, ITALY, JAPAN, REPUBLIC OF KOREA, LATVIA, LIBYA, MEXICO, NORWAY, POLAND, RUSSIA, SINGAPORE, SPAIN, SWEDEN, TAIWAN, TURKEY, UNITED KINGDOM, ROMANIA, ISRAEL, UKRAINE, YEMEN, BELARUS, and AUSTRALIA.
Lot, serial or model codes
SYSTEM ID:  0847160005 542120XR02  561548BIP  561881BP1  630856H3131 817465AR3131  985867INNOVA31 COR373301 EG1000VA01 LY1086VA01 M4166990  05448VAS01 VAD2121  YE1002VA01 YV0365   NOT LOCOLIZED S/N CHILLER 798080-21 796259-09 794637-08 795371-01 796133-12 796259-05 796464-08 796464-13 796464-15 796767-05 796941-03 796941-04 796941-07 796941-09 797178-05 797179-03 797179-08 797179-12 797449-03 797449-06 797632-02 797812-04 797812-09 797812-10 797812-12 767326-03 770355-02 772153-13 767746-05 785678-16 794637-11 799172-07 799347-03 799347-05 796260-01 783679-01 797178-03 797179-07 797179-09 797179-11 797812-14 797812-16 799347-02 798449-01 798449-02 798449-03
510(k) numbers
["K060259","K061163"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register