Luminex Corporation: ARIES SARS-CoV-2 Assay, REF: 50-10047

Recall number
Z-0134-2023
Recalling firm
Luminex Corporation
Product
ARIES SARS-CoV-2 Assay, REF: 50-10047
Status
Open, Classified
Initiated
2022-09-14
Posted
2022-10-26
Terminated
not on file
Reason
A part defect with the wash 1 syringe of an SARS-CoV-2 assay could lead to incomplete extraction thereby producing an invalid or false negative assay result.
Root cause
Packaging process control
Action
On 09/14/22, recall notices were emailed to customers who were asked to do the following: 1. Cease all use, and isolate or quarantine affected devices pending disposal. 2. If affected devices are identified, contact the recalling firm at support@luminexcorp.com to arrange for replacement and confirm the disposal of any remaining kits in your inventory per the package insert disposal guidelines. 3. Perform a retrospective review of results obtained using the affected SARS-CoV-2 Assay. If you suspect that you have encountered a false negative result with a patient sample, please contact the recalling firm at support@luminexcorp.com. 4. Distribute the notification to any individuals within your organization who need to be made aware of this recall. 5. If affected devices were further distributed, notify your customers of this recall. 6. Please complete and return the Acknowledgement & Receipt Form. 7. For those with complaints, questions, or concerns, please contact Global Technical Support Services at 1-877-785-2323 (U.S. and Canada) or +1-512-381 -4397 (Outside U.S. and Canada). On 12/05/22, additional recall notices were sent to customers notifying them that two additional lots were impacted.
Quantity
5,141 (24 Cassettes)
Distribution
Distribution US nationwide, Indonesia, and Hong Kong.
Lot, serial or model codes
Lot - Expiration: AB4723A - 05/25/2022, AB5041A - 05/25/2022, AB5060A - 06/01/2022, AB5061A - 06/06/2022, AB5081A - 06/07/2022, AB5139A - 09/10/2022, AB5141A - 09/16/2022, AB5263A - 09/27/2022, AB5282A - 10/05/2022, AB6640A - 06/8/2023, AB6720A - 06/13/2023
510(k) numbers
not on file
PMA numbers
not on file
Product code
QJR
Firm FEI
3002524000
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register