AmniSure International LLC: AmniSure ROM Test, Fetal Membrane Rupture Test for In-Vitro Diagnostic Use, Model: FMRT-1, AmniSure International LLC, Cambridge, MA 02138
- Recall number
- Z-0120-2008
- Recalling firm
- AmniSure International LLC
- Product
- AmniSure ROM Test, Fetal Membrane Rupture Test for In-Vitro Diagnostic Use, Model: FMRT-1, AmniSure International LLC, Cambridge, MA 02138
- Status
- Terminated
- Initiated
- 2007-10-03
- Posted
- 2008-01-19
- Terminated
- 2008-02-06
- Reason
- Weak true postive/false negative results: weak true positive or false negatives in ruptured fetal membranes may occur due to a "hook effect"
- Root cause
- Device Design
- Action
- AmniSure notified users by Medical Device Correction (Recall) letter dated October 1, 2007 issued via US Mail. The letter provides additional labeling instructions and a checklist for trouble shooting, including a diluation procedure that allows users to determine and solve the problem of the hook effect with this device.
- Quantity
- 319,976 units
- Distribution
- Worldwide, USA, Australia, Austria, Italy, Korea, Bulgaria, Russia, Spain, Kuwait, Switzerland, Luexemburg, Ireland, Denmark, Netherlands, Jordan, Israel, Chile, UK, Singapore, Basil, New Zealand, Turkey, France, Greece, and Equador
- Lot, serial or model codes
- Lot Number/ Exp. Date A7003, 01/2009. A7002, 11/2008, A7001, 10/2008, A6005, 04/2008, A6004, 01/2008, A6003, 09/2007, and A6002-3, 09/2007
- 510(k) numbers
- ["K030849"]
- PMA numbers
- not on file
- Product code
- JJX
- Firm FEI
- 3005345832
- Firm city
- Cambridge
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28