Carl Zeiss Meditec AG: Carl Zeiss Meditec AG, Goeschwitzer Strasse 51-52 07745 Jena, Germany, IOLMaster 700, REF 1932-169 SN 1185393 2017-09-01
- Recall number
- Z-0108-2019
- Recalling firm
- Carl Zeiss Meditec AG
- Product
- Carl Zeiss Meditec AG, Goeschwitzer Strasse 51-52 07745 Jena, Germany, IOLMaster 700, REF 1932-169 SN 1185393 2017-09-01
- Status
- Terminated
- Initiated
- 2018-06-25
- Posted
- not on file
- Terminated
- 2021-01-15
- Reason
- Calibration setting was incorrectly entered during manufacturing causing the optical coherence tomography (OCT) to obtain the incorrect two-dimensional images of ocular structures of the anterior and posterior segments of the eye.
- Root cause
- Employee error
- Action
- On 08/06/2018 the firm sent a Field Safety Corrective Action letter to customers via FedEx Priority Overnight service informing them that based on routine post market surveillance activities, we have information that this specific IOLMaster 700 device is incorrectly calibrated. The calibration value for axial length measurement was not entered correctly in the device settings during the manufacturing process. The transmission discrepancy leads to an error during the axial length measurement of approx. 0.25mm. A deviation in the axial length of this magnitude results in an error of Single unit. approx. 0.8 Diopter with regard to the intraocular lens. This in turn results in a refractive error of about 0.5 Diopter. The local firm's service organization will remove this device from the field. The device manufacturer will recalibrate the device to ensure it meets specifications. Hazard involved: The likelihood of a faulty entered calibration value for axial length measurement in the production process of IOLMaster 700 is considered rare. The probability of injury, i.e. reduced visual acuity due to the implantation of a wrong IOL power, is very likely for the affected device, because the axial length of the eye cannot be measured correctly. The device failure may only be recognized through post-surgical examination. Affected product: Quantity: 1 Product Name: IOLMaster 700 Product Number: 000000-1932-169 Device Serial Number: 1185393 Manufacturing Date: 2017-09-18 Actions to be taken: 1. Discontinue use of the ZEISS IOLMaster 700 device (Serial Number 1185393). Physicians are asked to review refractive results during usage of this device and to inform the firm about any hyperopic refractive surprises noticed. Any measurement and calculation data from this device should not be taken into consideration for further treatments. 2. Complete and return the attached Medical Device Recall Return Response Form to USFCA.meditec.us@zeiss.com no later than Jun
- Quantity
- 1
- Distribution
- US: CA (Mountain View and San Jose) A single system. No distributed OUS.
- Lot, serial or model codes
- Model: 700 Catalog Number: 000000-1932-169 Serial Number: 1185393 UDI: 04049471092080
- 510(k) numbers
- ["K170171"]
- PMA numbers
- not on file
- Product code
- HJO
- Firm FEI
- 3001362763
- Firm city
- Jena
- Firm state
- not on file
- Firm country
- Germany
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28