Skip to the content

FDA Medical Device Recalls (openFDA)

Trumpf Medical Systems, Inc.: Helion and Xenion surgical lights, 501/01 series KLM lights. The surgical lights are intended to locally illuminate an operating or examination area of the patients body with a high intensity light.

Recall number
Z-0077-2013
Recalling firm
Trumpf Medical Systems, Inc.
Product
Helion and Xenion surgical lights, 501/01 series KLM lights. The surgical lights are intended to locally illuminate an operating or examination area of the patients body with a high intensity light.
Status
Terminated
Initiated
2012-09-19
Posted
2012-10-18
Terminated
2014-01-21
Reason
Incidents regarding fractures of the front joint of the spring arm have been reported.
Root cause
Device Design
Action
Trumpf sent an "URGENT RECALL NOTICE" dated September 19, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Confirmation of Receipt Form was attached for customers to complete and return. Contact the firm at 888-474-9359 for questions regarding this notice.
Quantity
275 units
Distribution
Nationwide Distribution-USA (nationwide) including the states of AR, CA, CO, CT, FL, GA, IL, KS, MA, MI, MO, NC, NJ, NY, OH, SC, TX, UT, VT and WI.
Lot, serial or model codes
Ondal AC2000 serial range of the spring arms requiring rework: 0111 010 xxxxxx to 0111 066 xxxxx (010 to 066 = month 01/year 20000 to month 06/year 2006 xxxxxx = 5 digit serial number).
510(k) numbers
["K011693"]
PMA numbers
not on file
Product code
FSY
Firm FEI
3003184737
Firm city
Charleston
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register