FDA Medical Device Recalls (openFDA)
Trumpf Medical Systems, Inc.: Helion and Xenion surgical lights, 501/01 series KLM lights. The surgical lights are intended to locally illuminate an operating or examination area of the patients body with a high intensity light.
- Recall number
- Z-0077-2013
- Recalling firm
- Trumpf Medical Systems, Inc.
- Product
- Helion and Xenion surgical lights, 501/01 series KLM lights. The surgical lights are intended to locally illuminate an operating or examination area of the patients body with a high intensity light.
- Status
- Terminated
- Initiated
- 2012-09-19
- Posted
- 2012-10-18
- Terminated
- 2014-01-21
- Reason
- Incidents regarding fractures of the front joint of the spring arm have been reported.
- Root cause
- Device Design
- Action
- Trumpf sent an "URGENT RECALL NOTICE" dated September 19, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Confirmation of Receipt Form was attached for customers to complete and return. Contact the firm at 888-474-9359 for questions regarding this notice.
- Quantity
- 275 units
- Distribution
- Nationwide Distribution-USA (nationwide) including the states of AR, CA, CO, CT, FL, GA, IL, KS, MA, MI, MO, NC, NJ, NY, OH, SC, TX, UT, VT and WI.
- Lot, serial or model codes
- Ondal AC2000 serial range of the spring arms requiring rework: 0111 010 xxxxxx to 0111 066 xxxxx (010 to 066 = month 01/year 20000 to month 06/year 2006 xxxxxx = 5 digit serial number).
- 510(k) numbers
- ["K011693"]
- PMA numbers
- not on file
- Product code
- FSY
- Firm FEI
- 3003184737
- Firm city
- Charleston
- Firm state
- SC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28